Perforations of right heart chambers associated with electrophysiology catheters and temporary transvenous pacing leads.

Farid Aliyev1, Cengiz Celiker, Cengizhan Türkoğlu, Bilgehan Karadağ, Ahmet Yıldız.   

Abstract

OBJECTIVES: Perforation of heart chambers is one of the rare complications observed during electrophysiological studies and placement of pacemaker leads. In this study, we performed a retrospective evaluation of patients with catheter-related right heart perforation, aiming to determine its incidence and clinical course. STUDY
DESIGN: We reviewed cases with catheter-related cardiac perforations observed at our institution from June 2002 to November 2007.
RESULTS: During the study period, a total of 2,385 procedures were performed (1,287 electrophysiologic studies, 1,098 temporary nonballoon-floating pacing lead placements). Eight cardiac perforations were diagnosed, with the overall procedure-based and catheter-based incidences of 0.34% (8/2,385) and 0.14% (8/5,603), respectively. Three of these perforations were related to diagnostic electrophysiology catheters, and five were related to temporary (1 permanent) transvenous pacemaker leads. Seven perforations involved the right ventricle and one involved the right atrium. Three patients in whom right ventricular perforation was detected at a late stage died suddenly after pacemaker lead implantation. One patient underwent surgical exploration because of right atrial perforation. Two patients underwent coronary bypass operation and, in one of these patients, perforation was detected during surgery. Two patients were managed conservatively.
CONCLUSION: Although right ventricular perforations detected early have a relatively benign course, those detected late and right atrial perforations require emergent surgical exploration and may have catastrophic consequences.

Entities:  

Mesh:

Year:  2011        PMID: 21358226

Source DB:  PubMed          Journal:  Turk Kardiyol Dern Ars        ISSN: 1016-5169


  4 in total

1.  A 9-year-old boy with severe diphtherial infection and cardiac complications.

Authors:  Charles Henry Washington; Satja Issaranggoon na ayuthaya; Krit Makonkawkeyoon; Peninnah Oberdorfer
Journal:  BMJ Case Rep       Date:  2014-11-20

2.  Asymptomatic Right Ventricular Perforation by a Temporary Transvenous Pacing Lead in an Infant.

Authors:  Hsu-Ting Cheng; Yun-Ching Fu; Chung-Chi Wang; Sheng-Ching Chan; Ming-Chih Lin; Sheng-Ling Jan
Journal:  Acta Cardiol Sin       Date:  2013-07       Impact factor: 2.672

3.  Temporary Endocavitary Pacemakers and their Use and Misuse: the Least is Better.

Authors:  Antoine Kossaify
Journal:  Clin Med Insights Cardiol       Date:  2014-01-08

4.  Subclinical cardiac perforation by cardiac implantable electronic device leads detected by cardiac computed tomography.

Authors:  Yeong-Min Lim; Jae-Sun Uhm; Min Kim; In-Soo Kim; Moo-Nyun Jin; Hee Tae Yu; Tae-Hoon Kim; Hye-Jeong Lee; Young-Jin Kim; Boyoung Joung; Hui-Nam Pak; Moon-Hyoung Lee
Journal:  BMC Cardiovasc Disord       Date:  2021-07-19       Impact factor: 2.298

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.