Literature DB >> 2134030

Peripheral atherectomy: experimental results with a new device.

N Nakagawa1, A H Cragg, T P Smith, S K Landas, S C De Jong.   

Abstract

The size of present rotational atherectomy devices is limited in part by a tendency to produce vessel torsion. The authors designed and investigated a large-bore rotational atherectomy device for peripheral atherectomy in a single pass without significant torsion. A plaque was retrieved from 36 of 40 cadaveric iliac arteries. The mean plaque size was 8.4 x 3.9 mm, and the average number retrieved per artery was two. Thirty of 34 severely calcified arteries were treated successfully. Effluent study revealed no distal embolization; however, six perforations and four dissections occurred. Preliminary results suggest that a cutting surface with a relatively large diameter can be designed to be effective without producing vessel torsion. Changes in future designs will include added flexibility and expandable cutting surfaces to enhance safety and minimize entry diameter.

Mesh:

Year:  1990        PMID: 2134030     DOI: 10.1016/s1051-0443(90)72517-1

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  1 in total

Review 1.  Atherectomy-assisted versus percutaneous angioplasty interventions for treatment of symptomatic infra-inguinal peripheral arterial disease.

Authors:  Obai Abdullah; Jad Omran; Ashraf S Al-Dadah; Kul Aggarwal; Tariq Enezate
Journal:  Arch Med Sci Atheroscler Dis       Date:  2019-11-21
  1 in total