Literature DB >> 21339246

New prototype of femoral arterial SmartCannula with anterograde and retrograde flow.

Denis Berdajs1, Enrico Ferrari, Alexandre Michalis, Marco Burki, Christian W Pieterse, Judith Horisberger, Ludwig K von Segesser.   

Abstract

OBJECTIVES: Femoral artery cannulation is routinely used in circulatory support scenarios for cardiorespiratory support in patients with acute cardiac and/or pulmonary decompensation. During prolonged perfusion, this may cause acute ischemia of the leg and, in the worst case, even amputation. The aim of this experimental study was to test a newly designed arterial cannula allowing proximal and distal blood flow.
METHODS: Veno-arterial cardiopulmonary bypass was established in three calves (67.6 ± 5.1 kg). The venous line was secured by cannulation of the external jugular vein. The arterial line of each animal was secured by cannulation of iliac arteries on both legs. On one side, we used a modified SmartCannula (SmartCannula 18Fr 130 mm) and, on the other side, a standard rectilinear BioMedicus cannula system was used, 19 Fr for retrograde and an 8 Fr BioMedicus shunt for anterograde flow toward the leg. An ultrasonic flow probe was used to quantify the perfusion of the distal leg.
RESULTS: At 0.5, 1.0, 1.5, and 2.0 l/min, the anterograde leg blood flow increased steadily for each cannula, but remained higher in the self-expanding cannula. That is, the 8 F rectilinear cannula achieved a blood flow of 0.02 ± 0.01, 0.1 ± 0.04, 0.22 ± 0.09, and 0.21 ± 0.02 l/min, respectively, and the 18 F self-expanding cannula achieved 0.06 ± 0.02, 0.15 ± 0.03, 0.24 ± 0.07 and 0.36 ± 0.04 l/min.
CONCLUSION: The modified self-expanding cannula exhibited superior distal arterial flow compared to the routinely used rectilinear shunt system. This has a potential to reduce ischemic events of the inferior extremity in prolonged perfusion.

Entities:  

Mesh:

Year:  2011        PMID: 21339246     DOI: 10.1177/0267659111399953

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  3 in total

1.  Extracorporeal membrane oxygenation for accidental deep hypothermia-current challenges and future perspectives.

Authors:  Piotr Mazur; Sylweriusz Kosiński; Paweł Podsiadło; Anna Jarosz; Roman Przybylski; Radosław Litiwnowicz; Jacek Piątek; Janusz Konstanty-Kalandyk; Robert Gałązkowski; Tomasz Darocha
Journal:  Ann Cardiothorac Surg       Date:  2019-01

2.  Pressure and Flow Characteristics of a Novel Bidirectional Cannula for Cardiopulmonary Bypass.

Authors:  Yi Chen; Elli Tutungi; James McMillan; Sara M Tayeh; Jess K Underwood; Adam C Wells; Julian A Smith; Randall A Moshinsky
Journal:  Innovations (Phila)       Date:  2017 Nov/Dec

3.  A Phase 1 Study of a Novel Bidirectional Perfusion Cannula in Patients Undergoing Femoral Cannulation for Cardiac Surgery.

Authors:  Silvana F Marasco; Elli Tutungi; Shirley A Vallance; Andrew A Udy; Justin C Negri; Adam D Zimmet; David C McGiffin; Vincent A Pellegrino; Randall A Moshinsky
Journal:  Innovations (Phila)       Date:  2018 Mar/Apr