Literature DB >> 21334912

Autotransfusion system or integrated automatic suction device in minimized extracorporeal circulation: influence on coagulation and inflammatory response.

Hansjörg Jenni1, Julia Rheinberger, Martin Czerny, Erich Gygax, Robert Rieben, Eva Krähenbühl, Thierry Carrel, Mario Stalder.   

Abstract

OBJECTIVE: To measure surrogate markers of coagulation activation as well as of the systemic inflammatory response in patients undergoing primary elective coronary artery bypass grafting (CABG) using either the so-called Smart suction device or a continuous autotransfusion system (C.A.T.S.®).
METHODS: Fifty-eight patients being operated with a miniaturized circuit (minimal extracorporeal circuit, MECC) were prospectively randomized to using a so-called Smart suction device or a routine continuous autotransfusion system (C.A.T.S.®) for collection of mediastinal shed blood. The coagulation response was measured by thrombin-antithrombin complex (TAT) and D-dimer. The inflammatory response was measured by Interleukin 6 (IL-6) and complement factor 3a (C3a) at three different time points, before surgery, 2h after surgery, as well as 18 h after surgery.
RESULTS: No serious adverse cardiovascular event was observed. Serum levels of TAT significantly differed between both groups 2h after surgery (Smart suction 16.12 ± 13.51 μg l⁻¹ vs C.A.T.S® 9.83 ± 7.81 μg l⁻¹, p = 0.040) and returned to baseline values after 18 h in both groups. Serum levels of D-dimer showed a corresponding pattern with a peak 2h after surgery (Smart suction 1115 ± 1231 ng ml⁻¹ vs C.A.T.S.® 507 ± 604 ng ml⁻¹, p = 0.025). IL-6 levels also significantly differed between both groups 2h after surgery (Smart suction 186 ± 306 pg ml⁻¹ vs C.A.T.S.® 82 ± 71 pg ml⁻¹, p = 0.072). No significant changes in serum levels of C3a over time could be observed.
CONCLUSIONS: Despite no differences in the clinical course of patients with either Smart suction or C.A.T.S.® being observed, surrogate markers of coagulation and inflammation seem to be less pronounced in patients where cardiotomy blood is not being directly reinfused. As such, C.A.T.S.® should be preferred in routine CABG, as long as no extensive volume substitution is anticipated.
Copyright © 2011 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21334912     DOI: 10.1016/j.ejcts.2010.11.082

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  1 in total

1.  Is elimination of cardiotomy suction preferable in aortic valve replacement? Assessment of perioperative coagulation, fibrinolysis and inflammation.

Authors:  Akimasa Morisaki; Atsushi Nakahira; Yasuyuki Sasaki; Hidekazu Hirai; Yuko Okada; Shigefumi Suehiro; Toshihiko Shibata
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-05-31
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.