Biomechanical and histologic analysis in aortic endoprosthesis using fibrin glue.

Marcelo José Almeida1, Winston Bonetti Yoshida, Ludvig Hafner, Julio Sequeira, Juliana Henriques Dos Santos, Ana Paula Batista Masseno, José Bitu Moreno, Sílvia de Sá Lorena.   

Abstract

BACKGROUND: The absence of incorporation between endoprosthesis (EP) and the arterial wall may lead to device migration and endoleaks around the stent graft. Alternatives have been tested aiming to improve this incorporation. Fibrin glue is used in many operating procedures promoting adhesion and tissue regeneration; however, its use to improve EP incorporation by arteries is unknown.
OBJECTIVE: The objective of this study was to analyze dislodgement forces needed to extract the EPs implanted in pig aorta, compare different oversizing and fibrin glue injections, and to analyze histologic changes among groups.
METHODS: Straight EPs were implanted in the thoracic aorta of pigs using 10% oversizing plus fibrin glue in the interface between the EP and the artery (group 1), using 20% oversizing (group 2), and 10% oversizing (group 3). Fourteen days after the implant, the animals were killed to enable biomechanical analysis of the EP and to verify histologic changes of the aortic wall and its interface with the EP.
RESULTS: Group 1 showed a dislodgement force of 21.9 ± 5.3 Newton (N) being higher than the other groups and statistically significant when compared to group 3 (15.6 ± 3.6N), P = .003%. Group 2 had a higher dislodgement force and statistically more significant than group 3 (19.5 ± 7.8N). Histologic analysis showed tissue reaction with inflammatory cells and fibroblasts higher in group 1 and group 2 compared to group 3.
CONCLUSION: This study reports a large animal survival model of thoracic aortic stent graft placement by testing the impact of fibrin glue on EP incorporation. Compared to oversizing alone, fibrin glue placed between the stent graft and the arterial wall increases EP incorporation. Additional studies are needed to determine the potential utility of fibrin glue in the setting of human arterial endografts.
Copyright © 2011 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21334167     DOI: 10.1016/j.jvs.2010.11.112

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  3 in total

1.  Early histological changes in the porcine aortic media after thoracic stent-graft implantation.

Authors:  Johannes Scheumann; Claudia Heilmann; Friedhelm Beyersdorf; Matthias Siepe; Robert M Brenner; Dittmar Böckler; Randall B Griepp; Moritz S Bischoff
Journal:  J Endovasc Ther       Date:  2012-06       Impact factor: 3.487

2.  How to design the optimal self-expandable oesophageal metallic stents: 22 years of experience in 645 patients with malignant strictures.

Authors:  Han Kyu Na; Ho-Young Song; Jin Hyoung Kim; Jung-Hoon Park; Min Kyoung Kang; Jongjin Lee; Se Jin Oh
Journal:  Eur Radiol       Date:  2012-09-27       Impact factor: 5.315

3.  Evaluation of Local Tissue Reaction After the Application of a 3D Printed Novel Holdfast Device for Left Atrial Appendage Exclusion.

Authors:  Maciej Brzeziński; Aleksandra Sejda; Rafał Pęksa; Maciej Pawlak; Kamil Bury; Zbigniew Adamiak; Maciej Kowalik; Dariusz Jagielak; Krzysztof Bartus; Mateusz K Hołda; Radoslaw Litwinowicz; Jan Rogowski
Journal:  Ann Biomed Eng       Date:  2019-07-15       Impact factor: 3.934