Literature DB >> 2133336

Dynamic ultraviolet sterilization of different implant types.

A A Delgado1, N G Schaaf.   

Abstract

This paper investigates the use of the dynamic ultraviolet sterilization process with various dental implants, stainless steel orthopedic cortical bone screws, and polysulfone polymer healing caps. These biomaterials were inoculated with the spores of Bacillus subtilis and Bacillus stearothermophilus. They were then exposed to dynamic ultraviolet radiation in the chamber of a BUD Ultraviolet Device. Samples were incubated in trypticase soy broth at 37 degrees C and 56 degrees C, and they were subcultured onto an enriched agar medium. Results indicate that 16 seconds of dynamic ultraviolet radiation is effective in sterilizing these materials. This is significantly less time than other sterilization techniques presently used.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2133336

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804