Effect of limb lead electrodes location on ECG and localization of idiopathic outflow tract tachycardia: a prospective study.

Arash Arya1, Yan Huo, Fredrick Frogner, Ulrike Wetzel, Philipp Sommer, Thomas Gaspar, Sergio Richter, Christopher Piorkowski, Gerhard Hindricks.   

Abstract

BACKGROUND: Different kinds of the surface ECG limb electrode positions may affect the limb lead vector and therefore the accuracy of the 12-lead ECG in localization of outflow tract ventricular tachycardia (OTVT). This study was intended to evaluate and compare the accuracy of the standard and the modified 12-lead ECG for localization of OTVT using the current published criteria. METHODS AND
RESULTS: Twenty consecutive patients (10 men, mean age, 51.6 ± 13.4 years) with OT-VT were included. A standard ECG with the distal placement of the limb lead electrodes and a modified ECG with the limb electrodes placed on the torso were recorded during the OT-VT and were used for localization by 2 electrophysiologists who were blinded to the successful ablation site to compare the accuracy of the 2 ECGs. The R wave amplitude during OT-VT in lead I of the standard 12-lead ECG was significantly higher compared to the modified surface ECG (0.225 ± 0.145 mV vs 0.139 ± 0.111 mV, P = 0.032). The S wave in aVR during OT-VT was significantly more negative compared to the modified surface ECG (-0.682 ± 0.182 mV vs -0.527 ± 0.228 mV, P = 0.017). The rate of accurate localization of the successful ablation sites in the anterior versus posterior outflow tract by the 2 observers using standard ECG (70% and 80%) were higher compared to modified ECG (50% and 60%, P = 0.042).
CONCLUSION: The R wave amplitude in lead I and the depth of the S wave amplitude in lead aVR of the standard surface ECG during OT-VT is significantly larger compared to the modified surface ECG. As the QRS morphology of the OT-VT is usually the first clue to the possible site of successful ablation, the standard 12-lead ECG should be used for more accurate localization of the origin of the OT-VT.
© 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21332862     DOI: 10.1111/j.1540-8167.2011.02012.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  3 in total

1.  S-R difference in V1-V2 is a novel criterion for differentiating the left from right ventricular outflow tract arrhythmias.

Authors:  Onur Kaypakli; Hasan Koca; Durmus Yıldıray Sahin; Fadime Karataş; Suleyman Ozbicer; Mevlüt Koç
Journal:  Ann Noninvasive Electrocardiol       Date:  2017-12-10       Impact factor: 1.468

Review 2.  12-lead Holter electrocardiography. Review of the literature and clinical application update.

Authors:  Li Su; Stefan Borov; Bernhard Zrenner
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-06-19

3.  A High Precision Machine Learning-Enabled System for Predicting Idiopathic Ventricular Arrhythmia Origins.

Authors:  Jianwei Zheng; Guohua Fu; Daniele Struppa; Islam Abudayyeh; Tahmeed Contractor; Kyle Anderson; Huimin Chu; Cyril Rakovski
Journal:  Front Cardiovasc Med       Date:  2022-03-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.