Literature DB >> 21332676

Drusen detection in retro-mode imaging by a scanning laser ophthalmoscope.

Jennifer H Acton1, Robert P Cubbidge, Helen King, Paul Galsworthy, Jonathan M Gibson.   

Abstract

PURPOSE: The Nidek F-10 is a scanning laser ophthalmoscope that is capable of a novel fundus imaging technique, so-called 'retro-mode' imaging. The standard method of imaging drusen in age-related macular degeneration (AMD) is by fundus photography. The aim of the study was to assess drusen quantification using retro-mode imaging.
METHODS: Stereoscopic fundus photographs and retro-mode images were captured in 31 eyes of 20 patients with varying stages of AMD. Two experienced masked retinal graders independently assessed images for the number and size of drusen, using purpose-designed software. Drusen were further assessed in a subset of eight patients using optical coherence tomography (OCT) imaging.
RESULTS: Drusen observed by fundus photography (mean 33.5) were significantly fewer in number than subretinal deposits seen in retro-mode (mean 81.6; p < 0.001). The predominant deposit diameter was on average 5 μm smaller in retro-mode imaging than in fundus photography (p = 0.004). Agreement between graders for both types of imaging was substantial for number of deposits (weighted κ = 0.69) and moderate for size of deposits (weighted κ = 0.42). Retro-mode deposits corresponded to drusen on OCT imaging in all eight patients.
CONCLUSION: The subretinal deposits detected by retro-mode imaging were consistent with the appearance of drusen on OCT imaging; however, a larger longitudinal study would be required to confirm this finding. Retro-mode imaging detected significantly more deposits than conventional colour fundus photography. Retro-mode imaging provides a rapid non-invasive technique, useful in monitoring subtle changes and progression of AMD, which may be useful in monitoring the response of drusen to future therapeutic interventions.
© 2011 The Authors. Acta Ophthalmologica © 2011 Acta Ophthalmologica Scandinavica Foundation.

Entities:  

Mesh:

Year:  2011        PMID: 21332676     DOI: 10.1111/j.1755-3768.2011.02123.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  14 in total

1.  Retromode imaging: Review and perspectives.

Authors:  Won June Lee; Byung Ro Lee; Yong Un Shin
Journal:  Saudi J Ophthalmol       Date:  2014-02-20

2.  Automated drusen detection in dry age-related macular degeneration by multiple-depth, en face optical coherence tomography.

Authors:  Rui Zhao; Acner Camino; Jie Wang; Ahmed M Hagag; Yansha Lu; Steven T Bailey; Christina J Flaxel; Thomas S Hwang; David Huang; Dengwang Li; Yali Jia
Journal:  Biomed Opt Express       Date:  2017-10-17       Impact factor: 3.732

3.  The noninvasive retro-mode imaging of confocal scanning laser ophthalmoscopy in myopic maculopathy: a prospective observational study.

Authors:  Y Su; X Zhang; K Wu; Y Ji; C Zuo; M Li; F Wen
Journal:  Eye (Lond)       Date:  2014-06-13       Impact factor: 3.775

4.  In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy.

Authors:  Despina Kokona; Joël Jovanovic; Andreas Ebneter; Martin S Zinkernagel
Journal:  J Vis Exp       Date:  2017-11-11       Impact factor: 1.355

Review 5.  Cholesterol in the retina: the best is yet to come.

Authors:  Irina A Pikuleva; Christine A Curcio
Journal:  Prog Retin Eye Res       Date:  2014-04-04       Impact factor: 21.198

6.  Retro mode illumination for detecting and quantifying the area of geographic atrophy in non-neovascular age-related macular degeneration.

Authors:  Giulia Corradetti; Iksoo Byon; Federico Corvi; Mariano Cozzi; Giovanni Staurenghi; SriniVas R Sadda
Journal:  Eye (Lond)       Date:  2021-07-14       Impact factor: 4.456

7.  The noninvasive retro-mode imaging modality of confocal scanning laser ophthalmoscopy in polypoidal choroidal vasculopathy: a preliminary application.

Authors:  Renpan Zeng; Xiongze Zhang; Yu Su; Meng Li; Kunfang Wu; Feng Wen
Journal:  PLoS One       Date:  2013-09-18       Impact factor: 3.240

Review 8.  Clinical applications of fundus autofluorescence in retinal disease.

Authors:  Madeline Yung; Michael A Klufas; David Sarraf
Journal:  Int J Retina Vitreous       Date:  2016-04-08

9.  Retro-mode imaging and fundus autofluorescence with scanning laser ophthalmoscope of retinal dystrophies.

Authors:  Battaglia Parodi Maurizio; Iacono Pierluigi; Kontadakis Stelios; Vergallo Stefano; Cascavilla Marialucia; Zucchiatti Ilaria; Bandello Francesco
Journal:  BMC Ophthalmol       Date:  2012-05-15       Impact factor: 2.209

10.  Monte Carlo simulation of laser beams interaction with the human eye using Geant4.

Authors:  Diogo Tendeiro; Gonçalo Lopes; Pedro Vieira; José Paulo Santos
Journal:  Biomed Eng Online       Date:  2014-05-07       Impact factor: 2.819

View more