Abstract
The purpose of this morbidity and mortality case report is to discuss a characteristic of the Viabahn (W.L. Gore, Newark, DE) stent-graft deployment as well as a complication related to wire trauma during endovascular exclusion of a popliteal artery aneurysm. We encountered a phenomenon not described in the literature during deployment of a Viabahn stent graft. This necessitated further wire purchase, which resulted in perforation of a branch vessel within the calf. The rapid recognition of the clinical findings and prompt diagnosis of compartment syndrome was essential to early intervention, which prevented further complication and permanent sequelae.Entities:
Keywords: Popliteal aneurysm; compartment syndrome; fasciotomy; stent graft
Year: 2007 PMID: 21326474 PMCID: PMC3036317 DOI: 10.1055/s-2007-985740
Source DB: PubMed Journal: Semin Intervent Radiol ISSN: 0739-9529 Impact factor: 1.513