Introduction and clinical evaluation of servo-controlled applanation resonance tonometry.

Gauti Jóhannesson1, Per Hallberg, Anders Eklund, Christina Lindén.   

Abstract

PURPOSE: In recent years, Applanation Resonance Tonometry (ART) has been suggested for intraocular pressure (IOP) measurements. The manual version of ART (ART(manual) ) has been further developed, and to improve usability, an automatic servo-controlled prototype (ART(servo) ) has been proposed. The aim of this study was to assess the limits of agreement (LoA) of ART(manual) and ART(servo) as compared with the reference method, Goldmann Applanation Tonometry (GAT).
METHODS: This was a prospective single-centre study on 152 eyes from 77 subjects. It was designed according to International Standard Organization's (ISO) requirements for tonometers (ISO 2001). Intraocular pressure was measured six times/method in a standardized order. The ART technique has two available analysis procedures: a dynamic one that measures IOP during the indentation phase and a static one that causes a Goldmann-like measurement during two seconds of full applanation. The 95% LoA was defined as ±1.96 × standard deviation of difference against GAT.
RESULTS: Mean IOP for GAT was 19.1 mmHg (range: 10-37 mmHg). The 95% LoA of ART(manual) was ±4.5 mmHg for both dynamic and static analyses. The 95% LoA of ART(servo) was ±5.7 mmHg for dynamic and ±4.9 mmHg for static analyses.
CONCLUSIONS: This study confirms that the ART methodology is feasible. The further developed ART(manual) fulfilled the ISO standard with both the dynamic and the static analysis techniques. ART(servo) with static analysis was close to fulfilling the standard but failed to do so in the highest IOP range. ART(servo) has the potential to greatly improve usability if further development is completed.
© 2011 The Authors. Acta Ophthalmologica © 2011 Acta Ophthalmologica Scandinavica Foundation.

Mesh:

Year:  2011        PMID: 21323868     DOI: 10.1111/j.1755-3768.2011.02111.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  3 in total

1.  Age-dependency of ocular parameters: a cross sectional study of young and elderly healthy subjects.

Authors:  G Jóhannesson; P Hallberg; K Ambarki; A Eklund; C Lindén
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-08-13       Impact factor: 3.117

2.  Repeatability and reproducibility of applanation resonance tonometry: a cross-sectional study.

Authors:  Laura Ottobelli; Paolo Fogagnolo; Paolo Frezzotti; Stefano De Cillà; Elena Vallenzasca; Maurizio Digiuni; Ruggiero Paderni; Ilaria Motolese; Simone Alex Bagaglia; Eduardo Motolese; Luca Rossetti
Journal:  BMC Ophthalmol       Date:  2015-04-10       Impact factor: 2.209

Review 3.  Intracranial and Intraocular Pressure at the Lamina Cribrosa: Gradient Effects.

Authors:  Gauti Jóhannesson; Anders Eklund; Christina Lindén
Journal:  Curr Neurol Neurosci Rep       Date:  2018-04-12       Impact factor: 5.081

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.