Are the point-of-care diagnostics MULTIPLATE and ROTEM valid in the setting of high concentrations of heparin and its reversal with protamine?

Ralph Gertler1, Gunther Wiesner, Peter Tassani-Prell, Siegmund-Lorenz Braun, Klaus Martin.   

Abstract

OBJECTIVES: To evaluate the in vitro effects of high concentrations of heparin and its reversal with protamine on routine laboratory parameters as well as on modified thromboelastogram (ROTEM; TEM International, Munich, Germany) and impedance aggregometry (MULTIPLATE; Dynabyte, Munich, Germany).
DESIGN: An observational, nonrandomized in vitro study.
SETTING: A single-center, university hospital. PARTICIPANTS: Ten healthy volunteers.
INTERVENTIONS: Heparinization of whole blood to levels of 2, 4, 6, and 8 IU/mL of heparin and reversal with protamine. For MULTIPLATE measurements, heparin levels up to 20 IU/mL were tested.
MEASUREMENTS AND MAIN RESULTS: The present results show that the prothrombin time (PT) and fibrinogen measurements are altered significantly by heparin concentrations above 2 IU/mL. Protamine reversal also affected coagulation tests except for the fibrinogen. The INTEM test using the ROTEM system was influenced significantly by heparin concentrations of 2 IU/mL or higher, whereas EXTEM measurements remained stable up to 4 IU/mL. The findings for the FIBTEM test were stable up to 6 IU/mL but then declined to values less than 50% of baseline at 8 IU/mL. HEPTEM results remained valid under all concentrations of heparin tested. The effect of protamine on ROTEM was seen mainly in the INTEM and HEPTEM measurements. Heparin concentrations up to a level of 20 U/mL had no effect on MULTIPLATE measurements. Effects of protamine on MULTIPLATE became significant at heparin-to-protamine ratios below 1:1 and were more pronounced for adenosine diphosphate than for thrombin receptor-activated protein testing.
CONCLUSIONS: Neither fibrinogen (Clauss) nor derived fibrinogen or FIBTEM testing is valid in the setting of high concentrations of heparin unless antagonized by heparinase. Reversal of heparin with protamine worsens platelet function at all ratios as detected by aggregometry (MULTIPLATE) and thromboelastography (ROTEM), starting at a 1:1 ratio. Therefore, appropriate coagulation testing under cardiopulmonary bypass conditions should be selected carefully according to heparin levels. In particular, fibrinogen values are falsely low at heparin levels of 2 IU/mL and above. Therefore, newer algorithms promoting the correction of fibrinogen levels on cardiopulmonary bypass should be based on appropriate testing.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21315618     DOI: 10.1053/j.jvca.2010.11.020

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  12 in total

Review 1.  An introduction to point-of-care testing in extracorporeal circulation and LVADs.

Authors:  Rachel Sara Bercovitz
Journal:  Hematology Am Soc Hematol Educ Program       Date:  2018-11-30

2.  Enhanced Photoacoustic Detection of Heparin in Whole Blood via Melanin Nanocapsules Carrying Molecular Agents.

Authors:  Wonjun Yim; Kathryn Takemura; Jiajing Zhou; Jingcheng Zhou; Zhicheng Jin; Raina M Borum; Ming Xu; Yong Cheng; Tengyu He; William Penny; Bill R Miller; Jesse V Jokerst
Journal:  ACS Nano       Date:  2021-12-28       Impact factor: 18.027

3.  A Nanoscale Tool for Photoacoustic-Based Measurements of Clotting Time and Therapeutic Drug Monitoring of Heparin.

Authors:  Junxin Wang; Fang Chen; Santiago J Arconada-Alvarez; James Hartanto; Li-Peng Yap; Ryan Park; Fang Wang; Ivetta Vorobyova; Grant Dagliyan; Peter S Conti; Jesse V Jokerst
Journal:  Nano Lett       Date:  2016-09-28       Impact factor: 11.189

4.  Prediction of Post-Weaning Fibrinogen Status during Cardiopulmonary Bypass: An Observational Study in 110 Patients.

Authors:  Gabor Erdoes; Germaine Gerster; Giuseppe Colucci; Heiko Kaiser; Lorenzo Alberio; Balthasar Eberle
Journal:  PLoS One       Date:  2015-05-26       Impact factor: 3.240

5.  Predictive value of rotational thromboelastometry during cardiopulmonary bypass for thrombocytopenia and hypofibrinogenemia after weaning of cardiopulmonary bypass.

Authors:  Sung-Mi Ji; Sung-Hoon Kim; Jae-Sik Nam; Hye-Joo Yun; Jeong-Hyun Choi; Eun-Ho Lee; In-Cheol Choi
Journal:  Korean J Anesthesiol       Date:  2015-05-28

6.  The role of evidence-based algorithms for rotational thromboelastometry-guided bleeding management.

Authors:  Klaus Görlinger; Antonio Pérez-Ferrer; Daniel Dirkmann; Fuat Saner; Marc Maegele; Ángel Augusto Pérez Calatayud; Tae-Yop Kim
Journal:  Korean J Anesthesiol       Date:  2019-05-17

7.  Less is more: We are administering too much protamine in cardiac surgery.

Authors:  Francesco De Simone; Pasquale Nardelli; Margherita Licheri; Giovanna Frau; Martina Baiardo Redaelli; Fabrizio Monaco; Alberto Zangrillo; Giovanni Landoni
Journal:  Ann Card Anaesth       Date:  2021 Apr-Jun

8.  Time course of haemostatic effects of fibrinogen concentrate administration in aortic surgery.

Authors:  C Solomon; C Hagl; N Rahe-Meyer
Journal:  Br J Anaesth       Date:  2013-02-06       Impact factor: 9.166

9.  Impact of different dosage of protamine on heparin reversal during off-pump coronary artery bypass: a clinical study.

Authors:  Salvatore Suelzu; Andrea Cossu; Giovanni Pala; Michele Portoghese; Valeria Columbanu; Gabriele Sales; Luigi Solinas; Luca Brazzi
Journal:  Heart Lung Vessel       Date:  2015

Review 10.  Thromboelastography and Thromboelastometry in Assessment of Fibrinogen Deficiency and Prediction for Transfusion Requirement: A Descriptive Review.

Authors:  Henry T Peng; Bartolomeu Nascimento; Andrew Beckett
Journal:  Biomed Res Int       Date:  2018-11-25       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.