Comparison of oxygen saturation values obtained from fingers on physically restrained or unrestrained sides of the body.

Esra Akin Korhan1, Gülendam Hakverdioğlu Yönt, Leyla Khorshid.   

Abstract

OBJECTIVE: The aim of this study was to compare semiexperimentally the pulse oximetry values obtained from a finger on restrained or unrestrained sides of the body.
BACKGROUND: The pulse oximeter provides a noninvasive measurement of the oxygen saturation of hemoglobin in arterial blood. One of the procedures most frequently applied to patients in intensive care units is the application of physical restraint. Circulation problems are the most important complication in patients who are physically restrained. Evaluation of oxygen saturation from body parts in which circulation is impeded or has deteriorated can cause false results.
METHODS: The research sample consisted of 30 hospitalized patients who participated in the study voluntarily and who were concordant with the inclusion criteria of the study. Patient information and patient follow-up forms were used for data collection. Pulse oximetry values were measured simultaneously using OxiMax Nellcor finger sensors from fingers on the restrained and unrestrained sides of the body. Numeric and percentile distributions were used in evaluating the sociodemographic properties of patients.
RESULTS: A significant difference was found between the oxygen saturation values obtained from a finger of an arm that had been physically restrained and a finger of an arm that had not been physically restrained. The mean oxygen saturation value measured from a finger of an arm that had been physically restrained was found to be 93.40 (SD, 2.97), and the mean oxygen saturation value measured from a finger of an arm that had not been physically restrained was found to be 95.53 (SD, 2.38).
CONCLUSIONS: The results of this study indicate that nurses should use a finger of an arm that is not physically restrained when evaluating oxygen saturation values to evaluate them correctly.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21311247     DOI: 10.1097/NUR.0b013e31820aeff2

Source DB:  PubMed          Journal:  Clin Nurse Spec        ISSN: 0887-6274            Impact factor:   1.067


  1 in total

1.  Accuracy of pulse oximetry in detection of oxygen saturation in patients admitted to the intensive care unit of heart surgery: comparison of finger, toe, forehead and earlobe probes.

Authors:  Sohila Seifi; Alireza Khatony; Gholamreza Moradi; Alireza Abdi; Farid Najafi
Journal:  BMC Nurs       Date:  2018-04-17