A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer.

Jorge Jorge1, José M González-Méijome, António Queirós, Paulo Fernandes, José A Diaz-Rey.   

Abstract

PURPOSE: The aim of this study was to evaluate the accuracy of intraocular pressure (IOP) values from the new non-contact tonometer (NCT) Reichert R7 by comparing results with those from Goldmann applanation tonometry (GAT) and the Reichert Ocular Response Analyzer (ORA). Other ocular dimensions were assessed to evaluate their potential influence on the IOP values obtained.
METHODS: Ninety two right eyes from 92 adults aged between 21 and 59 years (mean 34.9 ± 11.7 years) were enrolled in this study. IOP was measured with R7, ORA and GAT. All measurements were taken between 14:00 and 16:00 in the afternoon. Corneal resistance factor (CRF) and corneal hysteresis were measured with the Reichert ORA. The spherical equivalent refractive error was obtained using an open field auto-refractor (WAM5500; Grand Seiko) and corneal curvature, anterior chamber depth, corneal diameter and axial length were assessed with an optical coherence biometer (IOL Master; Zeiss Meditec, CA, USA).
RESULTS: The mean values for IOP measurements were 15.20 ± 3.37 mmHg (R7), 13.49 ± 3.55 mmHg (GAT), 15.01 ± 3.38 mmHg (ORA IOPcc) and 14.44 ± 3.47 mmHg (ORA IOPg). With the exception of the CRF (ρ = 0.72 p < 0.001) the correlations between ocular parameters and IOP obtained with the R7 were neither statistically nor clinically significant.
CONCLUSIONS: The new NCT, R7 overestimated the IOP compared with GAT in normal, healthy eyes by about 1.7 mmHg on average (95% confidence range of approximately -2 to +6 mmHg). The measures provided by the R7 were significantly influenced by the stiffness of the corneal tissue as measured by the ORA CRF value but not by other dimensional parameters of the eye.
© 2011 The College of Optometrists.

Entities:  

Mesh:

Year:  2011        PMID: 21309804     DOI: 10.1111/j.1475-1313.2010.00817.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  3 in total

1.  Ocular surface displacement with and without contact lenses during non-contact tonometry.

Authors:  Ulfah Rimayanti; Yoshiaki Kiuchi; Shohei Uemura; Joji Takenaka; Hideki Mochizuki; Makoto Kaneko
Journal:  PLoS One       Date:  2014-04-29       Impact factor: 3.240

2.  Intraocular Pressure Based on Dynamic Bidirectional Applanation and Air-puff Tonometry: A Comparative Study.

Authors:  Ayse E Bahadir Kilavuzoglu; Cemile B Cosar; Ali Rc Celebi; Ugur E Al Parmak
Journal:  J Curr Glaucoma Pract       Date:  2019 May-Aug

3.  Comparison of Non-contact Tonometry and Goldmann Applanation Tonometry Measurements in Non-pathologic High Myopia.

Authors:  Peiyuan Wang; Yunhe Song; Fengbin Lin; Zhenyu Wang; Xinbo Gao; Weijing Cheng; Meiling Chen; Yuying Peng; Yuhong Liu; Xiulan Zhang; Shida Chen
Journal:  Front Med (Lausanne)       Date:  2022-03-03
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.