A novel instrument for logging nearwork distance.

Tsz-wing Leung1, Daniel I Flitcroft, Josh Wallman, Tat Hing Lee, Yongping Zheng, Carly Siu-yin Lam, Chea-su Kee.   

Abstract

PURPOSE: To validate a novel ultrasonic sensor for logging reading distances. In addition, this device was used to compare the habitual reading distances between low and high myopes.
METHODS: First, the stability and sensitivity of the ultrasonic device were determined by repeated measures using artificial targets. Then, thirty Hong Kong Chinese (20-30 years) were recruited, of whom fifteen were considered to be high myopes (mean ± S.D. = -8.7 ± 0.5 D) and 15 to be low to non-myopes (mean ± S.D. = -2.0 ± 0.2 D). Each subject read a newspaper with their habitual visual aid continuously for 10 min in two sessions at their preferred working distance(s). The reading distances were recorded continuously using a novel nearwork analyzer. The modal working distance was considered as the 'habitual' reading distance. In addition, habitual reading distance was reported orally by each subject.
RESULTS: The nearwork analyzer gave accurate and repeatable measurements over a range of distances and angles. Using this instrument, high myopes were found to have a significantly shorter reading distance than low myopes or non-myopes (mean ± S.D. = 35.9 ± 9.8 cm vs 50.9 ± 24.8 cm; two-sample t-test, p = 0.04, df = 18). The reading distances reported orally by the subjects were not correlated with those recorded by the nearwork analyzer.
CONCLUSIONS: The nearwork analyzer was found to be an effective tool for measuring nearwork reading distance in a small group of emmetropic and myopic adults over a 10 min interval. Differences between the reading distance between high myopes and low/non-myopes was detected by the device. Further study is needed to determine if a closer working distance is a cause or effect of myopia development.
© 2011 The College of Optometrists.

Entities:  

Mesh:

Year:  2011        PMID: 21309800     DOI: 10.1111/j.1475-1313.2010.00814.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  5 in total

1.  Validation of the Clouclip and utility in measuring viewing distance in adults.

Authors:  Khob R Bhandari; Lisa A Ostrin
Journal:  Ophthalmic Physiol Opt       Date:  2020-10-01       Impact factor: 3.117

2.  Effects of Progressive Addition Lens Wear on Digital Work in Pre-presbyopes.

Authors:  Chea-Su Kee; Tsz Wing Leung; Ka-Hung Kan; Christie Hang-I Lam
Journal:  Optom Vis Sci       Date:  2018-05       Impact factor: 1.973

3.  Measures of accommodative function in secondary school year 9 and year 13: a 4-year longitudinal study.

Authors:  Esther Mármol-Errasti; José Miguel Cárdenas-Rebollo; Antonio Rodán; Encarnación Pagán-Fernández; Laura C Jara-García; Catalina Palomo-Álvarez
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-07-29       Impact factor: 3.535

4.  Near Work Related Parameters and Myopia in Chinese Children: the Anyang Childhood Eye Study.

Authors:  Shi-Ming Li; Si-Yuan Li; Meng-Tian Kang; Yuehua Zhou; Luo-Ru Liu; He Li; Yi-Peng Wang; Si-Yan Zhan; Bamini Gopinath; Paul Mitchell; Ningli Wang
Journal:  PLoS One       Date:  2015-08-05       Impact factor: 3.240

5.  The diversified defocus profile of the near-work environment and myopia development.

Authors:  Kai Yip Choi; Angela Yuen-Ting Mok; Chi-Wai Do; Paul Hong Lee; Henry Ho-Lung Chan
Journal:  Ophthalmic Physiol Opt       Date:  2020-06-09       Impact factor: 3.117

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.