Narrowed beam width in newer ultrasound machines shortens measurements in the lateral direction: fetal measurement charts may be obsolete.

I Økland1, T G Bjåstad, T F Johansen, H K Gjessing, P Grøttum, S H Eik-Nes.   

Abstract

OBJECTIVES: Fetal ultrasound measurements are made in axial, lateral and oblique directions. Lateral resolution is influenced by the beam width of the ultrasound system. To improve lateral resolution and image quality, the beam width has been made narrower; consequently, measurements in the lateral direction are affected and apparently made shorter, approaching the true length. The aims of this study were to explore our database to reveal time-dependent shortening of ultrasound measurements made in the lateral direction, and to assess the extent of beam-width changes by comparing beam-width measurements made on old and new ultrasound machines.
METHODS: A total of 41,941 femur length measurements, collected during the time-period 1987-2005, were analyzed, with time as a covariate. Using three ultrasound machines from the 1990s and three newer machines from 2007, we performed 25 series of blinded beam-width measurements on a tissue-mimicking phantom, measuring at depths of 3-8 cm with a 5-MHz transducer.
RESULTS: Regression analysis showed time to be a significant covariate. At the same gestational age, femur length measurement was 1.15 (95% CI, 1.08-1.23) mm shorter in the time-period 1999-2005 than in the time-period 1987-1992. Overall, the beam width was 1.08 (95% CI, 0.50-1.65) mm narrower with the new machines than with the old machines.
CONCLUSIONS: Technical improvements in modern ultrasound machines that have reduced the beam width affect fetal measurements in the lateral direction. This has clinical implications and new measurement charts are needed.
Copyright © 2011 ISUOG. Published by John Wiley & Sons, Ltd.

Mesh:

Year:  2011        PMID: 21308840     DOI: 10.1002/uog.8954

Source DB:  PubMed          Journal:  Ultrasound Obstet Gynecol        ISSN: 0960-7692            Impact factor:   7.299


  7 in total

1.  Fetal growth velocity: the NICHD fetal growth studies.

Authors:  Katherine L Grantz; Sungduk Kim; William A Grobman; Roger Newman; John Owen; Daniel Skupski; Jagteshwar Grewal; Edward K Chien; Deborah A Wing; Ronald J Wapner; Angela C Ranzini; Michael P Nageotte; Stefanie N Hinkle; Sarah Pugh; Hanyun Li; Karin Fuchs; Mary Hediger; Germaine M Buck Louis; Paul S Albert
Journal:  Am J Obstet Gynecol       Date:  2018-05-24       Impact factor: 8.661

2.  Intrauterine growth discordance across gestation and birthweight discordance in dichorionic twins.

Authors:  Melissa M Amyx; Paul S Albert; Alaina M Bever; Stefanie N Hinkle; John Owen; William A Grobman; Roger B Newman; Edward K Chien; Robert E Gore-Langton; Germaine M Buck Louis; Katherine L Grantz
Journal:  Am J Obstet Gynecol       Date:  2019-08-24       Impact factor: 8.661

3.  A comparison of four fetal biometry growth charts within an Australian obstetric population.

Authors:  Candice Dry; Michelle K Pedretti; Elizabeth Nathan; Jan E Dickinson
Journal:  Australas J Ultrasound Med       Date:  2022-02-24

4.  A survey of current practice in reporting third trimester fetal biometry and Doppler in Australia and New Zealand.

Authors:  Debra Paoletti; Lillian Smyth; Susan Westerway; Jon Hyett; Ritu Mogra; Stephen Haslett; Michael Peek
Journal:  Australas J Ultrasound Med       Date:  2021-08-26

5.  Dichorionic twin trajectories: the NICHD Fetal Growth Studies.

Authors:  Katherine L Grantz; Jagteshwar Grewal; Paul S Albert; Ronald Wapner; Mary E D'Alton; Anthony Sciscione; William A Grobman; Deborah A Wing; John Owen; Roger B Newman; Edward K Chien; Robert E Gore-Langton; Sungduk Kim; Cuilin Zhang; Germaine M Buck Louis; Mary L Hediger
Journal:  Am J Obstet Gynecol       Date:  2016-04-30       Impact factor: 8.661

6.  INTERGROWTH-21st - Time to standardise fetal measurement in Australia.

Authors:  Susan Campbell Westerway; Aris T Papageorghiou; Jane Hirst; Fabricio da Silva Costa; Jon Hyett; Susan P Walker
Journal:  Australas J Ultrasound Med       Date:  2015-12-31

7.  Fetal biometry assessment with Intergrowth 21st's and Salomon's equations in rural Burkina Faso.

Authors:  Biébo Bihoun; Serge Henri Zango; Maminata Traoré-Coulibaly; Innocent Valea; Raffaella Ravinetto; Jean-Pierre Van Geertruyden; Umberto D'Alessandro; Halidou Tinto; Annie Robert
Journal:  BMC Pregnancy Childbirth       Date:  2020-08-26       Impact factor: 3.007

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.