An eight-channel transmit/receive multipurpose coil for musculoskeletal MR imaging at 7 T.

Oliver Kraff1, Andreas K Bitz, Philipp Dammann, Susanne C Ladd, Mark E Ladd, Harald H Quick.   

Abstract

PURPOSE: MRI plays a leading diagnostic role in assessing the musculoskeletal (MSK) system and is well established for most questions at clinically used field strengths (up to 3 T). However, there are still limitations in imaging early stages of cartilage degeneration, very fine tendons and ligaments, or in locating nerve lesions, for example. 7 T MRI of the knee has already received increasing attention in the current published literature, but there is a strong need to develop new radiofrequency (RF) coils to assess more regions of the MSK system. In this work, an eight-channel transmit/receive RF array was built as a multipurpose coil for imaging some of the thus far neglected regions. An extensive coil characterization protocol and first in vivo results of the human wrist, shoulder, elbow, knee, and ankle imaged at 7 T will be presented.
METHODS: Eight surface loop coils with a dimension of 6 x 7 cm2 were machined from FR4 circuit board material. To facilitate easy positioning, two coil clusters, each with four loop elements, were combined to one RF transmit/receive array. An overlapped and shifted arrangement of the coil elements was chosen to reduce the mutual inductance between neighboring coils. A phantom made of body-simulating liquid was used for tuning and matching on the bench. Afterward, the S-parameters were verified on a human wrist, elbow, and shoulder. For safety validation, a detailed compliance test was performed including full wave simulations of the RF field distribution and the corresponding specific absorption rate (SAR) for all joints. In vivo images of four volunteers were assessed with gradient echo and spin echo sequences modified to obtain optimal image contrast, full anatomic coverage, and the highest spatial resolution within a reasonable acquisition time. The performance of the RF coil was additionally evaluated by in vivo B1 mapping.
RESULTS: A comparison of B1 per unit power, flip angle distribution, and anatomic images showed a fairly homogeneous excitation for the smaller joints (elbow, wrist, and ankle), while for the larger joints, the shoulder and especially the knee, B1 inhomogeneities and limited penetration depth were more pronounced. However, the greater part of the shoulder joint could be imaged. In vivo images rendered very fine anatomic details such as fascicles of the median nerve and the branching of the nerve bundles. High-resolution images of cartilage, labrum, and tendons could be acquired. Additionally, turbo spin echo (TSE) and inversion recovery sequences performed very well.
CONCLUSIONS: This study demonstrates that the concept of two four-channel transmit/receive RF arrays can be used as a multipurpose coil for high-resolution in vivo MR imaging of the musculoskeletal system at 7 T. Not only gradient echo but also typical clinical and SAR-intensive sequences such as STIR and TSE performed well. Imaging of small structures and peripheral nerves could in particular benefit from this technique.

Entities:  

Mesh:

Year:  2010        PMID: 21302794     DOI: 10.1118/1.3517176

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Bilateral hip imaging at 7 Tesla using a multi-channel transmit technology: initial results presenting anatomical detail in healthy volunteers and pathological changes in patients with avascular necrosis of the femoral head.

Authors:  J M Theysohn; O Kraff; S Orzada; N Theysohn; T Classen; S Landgraeber; M E Ladd; T C Lauenstein
Journal:  Skeletal Radiol       Date:  2013-08-18       Impact factor: 2.199

2.  Integrated parallel reception, excitation, and shimming (iPRES).

Authors:  Hui Han; Allen W Song; Trong-Kha Truong
Journal:  Magn Reson Med       Date:  2013-04-29       Impact factor: 4.668

3.  Ultra-high-field RF coil development for evaluating upper extremity imaging applications.

Authors:  Shailesh B Raval; Tiejun Zhao; Narayanan Krishnamurthy; Tales Santini; Cynthia Britton; Vijay S Gorantla; Tamer S Ibrahim
Journal:  NMR Biomed       Date:  2016-11-03       Impact factor: 4.044

4.  Ultra-high field upper extremity peripheral nerve and non-contrast enhanced vascular imaging.

Authors:  Shailesh B Raval; Cynthia A Britton; Tiejun Zhao; Narayanan Krishnamurthy; Tales Santini; Vijay S Gorantla; Tamer S Ibrahim
Journal:  PLoS One       Date:  2017-06-29       Impact factor: 3.240

5.  Modular transmit/receive arrays using very-high permittivity dielectric resonator antennas.

Authors:  Thomas P A O'Reilly; Thomas Ruytenberg; Andrew G Webb
Journal:  Magn Reson Med       Date:  2017-06-20       Impact factor: 4.668

6.  7-T clinical MRI of the shoulder in patients with suspected lesions of the rotator cuff.

Authors:  Andrea Lazik-Palm; Oliver Kraff; Stefan H G Rietsch; Mark E Ladd; Michael Kamminga; Sascha Beck; Harald H Quick; Jens M Theysohn
Journal:  Eur Radiol Exp       Date:  2020-02-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.