Finite element simulations of a focal knee resurfacing implant applied to localized cartilage defects in a sheep model.

Krishnagoud Manda1, Leif Ryd, Anders Eriksson.   

Abstract

Articular resurfacing metal implants have recently been tested in animal models to treat full thickness localized articular cartilage defects, showing promising results. However, the mechanical behavior of cartilage surrounding the metal implant has not been studied yet as it is technically challenging to measure in vivo contact areas, pressures, stresses and deformations from the metal implant. Therefore, we implemented a detailed numerical finite element model by approximating one of the condyles of the sheep tibiofemoral joint and created a defect of specific size to accommodate the implant. Using this model, the mechanical behavior of the surrounding of metal implant was studied. The model showed that the metal implant plays a significant role in the force transmission. Two types of profiles were investigated for metal implant. An implant with a double-curved profile, i.e., a profile fully congruent with the articular surfaces in the knee, gives lower contact pressures and stresses at the rim of the defect than the implant with unicurved spherical profile. The implant should be placed at a certain distance into the cartilage to avoid damage to opposing biological surface. Too deep positions, however, lead to high shear stresses in the cartilage edges around the implant. Mechanical sealing was achieved with a wedge shape of the implant, also useful for biochemical sealing of cartilage edges at the defect.
Copyright © 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21300358     DOI: 10.1016/j.jbiomech.2010.12.026

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  11 in total

1.  Cartilage Health in Knees Treated with Metal Resurfacing Implants or Untreated Focal Cartilage Lesions: A Preclinical Study in Sheep.

Authors:  Nicolas Martinez-Carranza; Kjell Hultenby; Anne Sofie Lagerstedt; Peter Schupbach; Hans E Berg
Journal:  Cartilage       Date:  2017-07-13       Impact factor: 4.634

2.  Focal metallic inlay resurfacing prosthesis for the treatment of localized cartilage defects of the femoral condyles: a systematic review of clinical studies.

Authors:  Andreas Fuchs; Helge Eberbach; Kaywan Izadpanah; Gerrit Bode; Norbert P Südkamp; Matthias J Feucht
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-09-16       Impact factor: 4.342

Review 3.  Advances in the Application of Three-dimensional Printing for the Clinical Treatment of Osteoarticular Defects.

Authors:  Xiao-Ze Fan; Ming-Ze Du; Chen Jiao; Qin-Wei Guo; Dong Jiang
Journal:  Curr Med Sci       Date:  2022-04-22

4.  Biphasic finite element contact analysis of the knee joint using an augmented Lagrangian method.

Authors:  Hongqiang Guo; Suzanne A Maher; Robert L Spilker
Journal:  Med Eng Phys       Date:  2013-03-15       Impact factor: 2.242

5.  Comparison between FEBio and Abaqus for biphasic contact problems.

Authors:  Qingen Meng; Zhongmin Jin; John Fisher; Ruth Wilcox
Journal:  Proc Inst Mech Eng H       Date:  2013-06-26       Impact factor: 1.617

6.  Cartilage defect location and stiffness predispose the tibiofemoral joint to aberrant loading conditions during stance phase of gait.

Authors:  Lianne Zevenbergen; Colin R Smith; Sam Van Rossom; Darryl G Thelen; Nele Famaey; Jos Vander Sloten; Ilse Jonkers
Journal:  PLoS One       Date:  2018-10-16       Impact factor: 3.240

7.  Effects of Loading Conditions on Articular Cartilage in a Metal-on-Cartilage Pairing.

Authors:  Christoph Stotter; Bojana Stojanović; Christoph Bauer; Manel Rodríguez Ripoll; Friedrich Franek; Thomas Klestil; Stefan Nehrer
Journal:  J Orthop Res       Date:  2019-08-06       Impact factor: 3.494

8.  The influence of size, clearance, cartilage properties, thickness and hemiarthroplasty on the contact mechanics of the hip joint with biphasic layers.

Authors:  Junyan Li; Todd D Stewart; Zhongmin Jin; Ruth K Wilcox; John Fisher
Journal:  J Biomech       Date:  2013-05-09       Impact factor: 2.712

9.  Measurement of the migration of a focal knee resurfacing implant with radiostereometry.

Authors:  Olof Sköldenberg; Thomas Eisler; André Stark; Olle Muren; Nicolas Martinez-Carranza; Leif Ryd
Journal:  Acta Orthop       Date:  2013-11-29       Impact factor: 3.717

10.  The effect of distance between holes on the structural stability of subchondral bone in microfracture surgery: a finite element model study.

Authors:  Xiang Yun Yin; Do Young Park; Young Jick Kim; Hye Jung Ahn; Seung-Hyun Yoo; Byoung-Hyun Min
Journal:  BMC Musculoskelet Disord       Date:  2020-08-18       Impact factor: 2.362