Mauro Bozzola1, Michel Colle, Maria Halldin-Stenlid, Sylvain Larroque, Monia Zignani.
Abstract
BACKGROUND: Accurately monitoring adherence to treatment with recombinant human growth hormone (r-hGH) enables appropriate intervention in cases of poor adherence. The electronic r-hGH auto-injector, easypod™, automatically records the patient's adherence to treatment. This study evaluated adherence to treatment of children who started using the auto-injector and assessed opinions about the device.Entities:
Year: 2011 PMID: 21294891 PMCID: PMC3045978 DOI: 10.1186/1472-6823-11-4
Source DB: PubMed Journal: BMC Endocr Disord ISSN: 1472-6823 Impact factor: 2.763
Survey questions evaluating acceptability of the electronic auto-injector
| Question | Response options | ||||
|---|---|---|---|---|---|
| How easy do you find the preparation steps of easypod™? | 1 | 2 | 3 | 4 | 5 |
| How long does it normally take you to prepare your growth hormone injection with easypod™? | 1 | 2 | 3 | 4 | 5 |
| How long does it normally take you to inject Saizen using easypod™? | 1 | 2 | 3 | 4 | 5 |
| For each injection site, which setting for the needle insertion depth do you prefer? | |||||
| Thighs | Low | Medium | High | Not applicable | |
| Belly | Low | Medium | High | Not applicable | |
| Buttocks | Low | Medium | High | Not applicable | |
| Arms | Low | Medium | High | Not applicable | |
| Which setting for the injection speed do you prefer? | Low | Medium | High | ||
| Did the injection with easypod™ hurt? | Yes | No | |||
| If yes, did the injection hurt as much as you thought it would? | 1 | 2 | 3 | 4 | 5 |
| Has easypod™ been easy to use? | 1 | 2 | 3 | 4 | 5 |
| Did you feel comfortable using easypod™? | 1 | 2 | 3 | 4 | 5 |
| If you had a choice, would you continue to use easypod™? | Yes | No | |||
| Do you think easypod™ is easy to hold/grip? | 1 | 2 | 3 | 4 | 5 |
| Do you think easypod™ is compact/portable? | 1 | 2 | 3 | 4 | 5 |
| Do you think easypod™ has an attractive/nice design? | 1 | 2 | 3 | 4 | 5 |
| Do you think it is easy to change the cartridge on easypod™? | 1 | 2 | 3 | 4 | 5 |
| Do you think it is easy to change the needle on easypod™? | 1 | 2 | 3 | 4 | 5 |
| Do you think injection using the easypod™ is quiet? | 1 | 2 | 3 | 4 | 5 |
Figure 1Geographic distribution of survey participants in the evaluable population.
Characteristics of the evaluable survey population
| Total | Treatment-naïve | Treatment-experienced | |
|---|---|---|---|
| Boys, n (%)a | 462 (56.1) | 358 (59.6) | 104 (46.9) |
| Age, median (range), years | 11 (1-18) | 11 (1-18) | 11 (2-18) |
| Indication, n (%)b | |||
| GHD | 543 (66.4) | 407 (68.1) | 136 (61.8) |
| TS | 80 (9.8) | 46 (7.7) | 34 (15.5) |
| CRF | 14 (1.7) | 10 (1.7) | 4 (1.8) |
| SGA | 125 (15.3) | 98 (16.4) | 27 (12.3) |
| Other | 56 (6.8) | 37 (6.2) | 19 (8.6) |
| Administration of injectionsc | |||
| Child | 315 (38.7) | 224 (37.5) | 91 (41.7) |
| Parent | 430 (52.8) | 317 (53.1) | 113 (51.8) |
| Other | 7 (0.86) | 6 (1.01) | 1 (0.46) |
| Child and parent | 59 (7.2) | 46 (7.7) | 13 (6.0) |
| Parent and other | 4 (0.49) | 4 (0.67) | 0 |
| Number of scheduled injections per weekd | |||
| 6 | 234 (29.3) | 177 (30.2) | 57 (26.9) |
| 7 | 564 (70.7) | 409 (69.8) | 155 (73.1) |
aOne child missing from the treatment-experienced group.
bSix children missing: three each from the treatment-naïve and treatment-experienced groups.
CNine children missing: four from the treatment-naïve group and five from the treatment-experienced group.
dTwenty-six children missing: fifteen from the treatment-naïve group and eleven from the treatment-experienced group.
CRF -- chronic renal failure; GHD -- growth hormone deficiency; SGA -- small for gestational age; TS -- Turner syndrome.
Adherence to treatment ( > 92% of injections taken)
| Total | Treatment-naïve | Treatment-experienced | |
|---|---|---|---|
| Overall, n (%)* | 649/772 (84.1) | 484/561 (86.3) | 165/211 (78.2) |
| Overall, n (%) | 597/682 (87.5) | 445/496 (89.7) | 152/186 (81.7) |
| Overall, n (%)* | 677/790 (85.7) | 497/573 (86.7) | 180/217 (83.0) |
| Overall, n (%) | 607/673 (90.2) | 447/489 (91.4) | 160/184 (87.0) |
| Month 1, n (%) | 617/682 (90.5) | 458/496 (92.3) | 159/186 (85.5) |
| Month 2, n (%) | 594/682 (87.1) | 441/496 (88.9) | 153/186 (82.3) |
| Month 3, n (%) | 606/682 (88.9) | 449/496 (90.5) | 157/186 (84.4) |
(a) Adherence as recorded and reported by children/parents overall over the 3-month survey. (b) Adherence as recorded by easypod™ over each month of the 3-month survey
*Overall adherence calculated by imputation of missing period(s) using non-missing period(s)
Figure 2Missed injections during each month of the survey according to the downloaded dose history (after imputation for missing data).
Figure 3Recorded adherence according to country (after imputation for missing data using non-missing period(s)).
Most common reasons given for missed injections
| Total | Treatment-naïve | Treatment-experienced | |
|---|---|---|---|
| Forgot injection | 180 (43.7) | 135 (45.6) | 45 (38.8) |
| Device not working | 75 (18.2) | 54 (18.2) | 21 (18.1) |
| Short vacations, slept away from home | 52 (12.6) | 31 (10.5) | 21 (18.1) |
| Ran out of cartridges/needles | 53 (12.9) | 43 (14.5) | 10 (8.6) |
| Forgot drug/easypod™ | 32 (7.8) | 25 (8.5) | 7 (6.0) |
| Sick | 35 (8.5) | 23 (7.8) | 12 (10.3) |
| Afraid of using easypod™ | 12 (2.9) | 5 (1.7) | 7 (6.0) |
| Tired of injections | 18 (4.4) | 7 (2.4) | 11 (9.5) |
| Reconstitution issues | 13 (3.2) | 11 (3.7) | 2 (1.7) |
| Lack of caregiver to provide injection | 5 (1.2) | 3 (1.0) | 2 (1.7) |
| Too complicated/forgot how to use device | 4 (0.97) | 2 (0.68) | 2 (1.7) |
| Other | 11 (2.7) | 6 (2.0) | 5 (4.3) |
Data values represent number (percentage) of respondents who gave the indicated response as a reason for missing injections
Acceptance of the electronic auto-injector
| Number | Most | → | Least | ||||
|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | |||
| Device preparation | Ease | 805 | 53.5 | 28.9 | 12.6 | 2.7 | 2.2 |
| Duration | 806 | 53.2 | 31.6 | 12.7 | 0.99 | 1.5 | |
| Device use | Ease | 779 | 62.6 | 29.8 | 5.8 | 1.0 | 0.77 |
| Duration | 805 | 53.0 | 31.9 | 13.3 | 1.2 | 0.50 | |
| Comfort | 804 | 57.5 | 31.6 | 7.6 | 1.7 | 1.6 | |
| Device features | Grip | 813 | 59.7 | 28.2 | 9.0 | 2.1 | 1.1 |
| Size | 803 | 52.4 | 26.0 | 15.6 | 3.6 | 2.4 | |
| Attractiveness | 807 | 56.6 | 28.3 | 12.8 | 1.5 | 0.87 | |
| Ease of cartridge change | 801 | 58.8 | 27.0 | 10.4 | 2.9 | 1.0 | |
| Ease of needle change | 809 | 56.9 | 26.1 | 12.2 | 3.7 | 1.1 | |
| Noise | 807 | 51.7 | 32.3 | 12.1 | 2.6 | 1.2 | |
Figures indicate the percentage of survey participants giving the indicated rating (scale of 1-5) in response to specific questions about the electronic auto-injector.