Longer storage of dialyzers increases elution of poly(N-vinyl-2-pyrrolidone) from polysulfone-group dialysis membranes.

Koki Namekawa1, Ami Kaneko, Kiyotaka Sakai, Satoru Kunikata, Masato Matsuda.   

Abstract

The objective of this study was to evaluate the effect of protracted storage of dialyzers on the amount of poly(N-vinyl-2-pyrrolidone) (PVP) eluted from polysulfone-group dialysis membranes. We tested five dialysis membranes: APS-15SA (Asahi Kasei Kuraray, wet), CX-1.6U (Toray, moist), FX140 (Fresenius, dry), PES-15Sα (Nipro, dry), and FDX-150GW (Nikkiso, wet). Each dialyzer was stored for 1, 3, 14, and 18 months after sterilization. The dialysis-fluid side compartment was primed with reverse osmosis (RO) water at 500 mL/min for 5 min at 310 K. The blood side compartment was primed with RO water at 200 mL/min for 5 min at 310 K. Finally, 1 L RO water was circulated through the blood side compartment at 200 mL/min for 4 h at 310 K. Eluted PVP was determined by use of the iodine method, using 0.02 N: iodine solution. PVP was mainly eluted from wet-type dialyzers during priming. Thus, the standard 5 min priming of the wet-type dialyzer according to the maker manual inhibits PVP elution during circulation. PVP was eluted in the dialysis-fluid side of the moist-type dialyzer during priming but no PVP was eluted in the blood side. PVP was mainly eluted from dry-type dialyzers during circulation. We recommend more than the standard 5 min priming, particularly for dry-type dialyzers stored for protracted periods, because 5 min insufficient to inhibit PVP elution during circulation.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21286768     DOI: 10.1007/s10047-011-0552-1

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  6 in total

1.  Polysulphone membrane mimicking human glomerular basement membrane.

Authors:  E Streicher; H Schneider
Journal:  Lancet       Date:  1983-11-12       Impact factor: 79.321

2.  Recalcitrance of poly(vinylpyrrolidone): evidence through matrix-assisted laser desorption-ionization time-of-flight mass spectrometry.

Authors:  S Trimpin; P Eichhorn; H J Räder; K Müllen; T P Knepper
Journal:  J Chromatogr A       Date:  2001-12-14       Impact factor: 4.759

3.  Nanotechnological characterization of human serum albumin adsorption on wet synthetic polymer dialysis membrane surfaces.

Authors:  Koki Namekawa; Makoto Fukuda; Masato Matsuda; Yutaka Yagi; Ken-ichiro Yamamoto; Kiyotaka Sakai
Journal:  ASAIO J       Date:  2009 May-Jun       Impact factor: 2.872

4.  [Identification and determination of polyvinylpyrrolidone(PVP) as well as determination of active substances in PVP-containing drug preparations].

Authors:  K Müller
Journal:  Pharm Acta Helv       Date:  1968 Jan-Feb

5.  Effects of fluid flow on elution of hydrophilic modifier from dialysis membrane surfaces.

Authors:  Masato Matsuda; Mika Sato; Hiroki Sakata; Takahisa Ogawa; Ken-ichiro Yamamoto; Taiji Yakushiji; Makoto Fukuda; Takehiro Miyasaka; Kiyotaka Sakai
Journal:  J Artif Organs       Date:  2008-10-05       Impact factor: 1.731

6.  Nanoscopic behavior of polyvinylpyrrolidone particles on polysulfone/polyvinylpyrrolidone film.

Authors:  Masayo Hayama; Ken-ichiro Yamamoto; Fukashi Kohori; Tsutomu Uesaka; Yoshiyuki Ueno; Hiroyuki Sugaya; Ichiro Itagaki; Kiyotaka Sakai
Journal:  Biomaterials       Date:  2004-03       Impact factor: 12.479

  6 in total
  3 in total

Review 1.  Journal of Artificial Organs 2011: the year in review.

Authors:  Y Sawa; E Tatsumi; A Funakubo; T Horiuchi; K Iwasaki; A Kishida; T Masuzawa; K Matsuda; A Myoui; M Nishimura; T Nishimura; S Tokunaga; Y Tomizawa; T Tomo; T Tsukiya; T Yamaoka
Journal:  J Artif Organs       Date:  2012-02-29       Impact factor: 1.731

2.  Poly(N-vinyl-2-pyrrolidone) elution from polysulfone dialysis membranes by varying solvent and wall shear stress.

Authors:  Koki Namekawa; Masato Matsuda; Makoto Fukuda; Ami Kaneko; Kiyotaka Sakai
Journal:  J Artif Organs       Date:  2012-02-05       Impact factor: 1.731

3.  A porcine model of hemodialyzer reactions: roles of complement activation and rinsing back of extracorporeal blood.

Authors:  Ákos Pethő; Dorothea Piecha; Tamás Mészáros; Rudolf Urbanics; Christoph Moore; Bernard Canaud; László Rosivall; Tom Eirik Mollnes; Sonja Steppan; Gábor Szénási; János Szebeni; László Dézsi
Journal:  Ren Fail       Date:  2021-12       Impact factor: 2.606

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.