Radiation exposure of the anesthesiologist in the neurointerventional suite.

Zirka H Anastasian1, Dorothea Strozyk, Philip M Meyers, Shuang Wang, Mitchell F Berman.   

Abstract

BACKGROUND: Scatter radiation during interventional radiology procedures can produce cataracts in participating medical personnel. Standard safety equipment for the radiologist includes eye protection. The typical configuration of fluoroscopy equipment directs radiation scatter away from the radiologist and toward the anesthesiologist. This study analyzed facial radiation exposure of the anesthesiologist during interventional neuroradiology procedures.
METHODS: Radiation exposure to the forehead of the anesthesiologist and radiologist was measured during 31 adult neuroradiologic procedures involving the head or neck. Variables hypothesized to affect anesthesiologist exposure were recorded for each procedure. These included total radiation emitted by fluoroscopic equipment, radiologist exposure, number of pharmacologic interventions performed by the anesthesiologist, and other variables.
RESULTS: Radiation exposure to the anesthesiologist's face averaged 6.5 ± 5.4 μSv per interventional procedure. This exposure was more than 6-fold greater (P < 0.0005) than for noninterventional angiographic procedures (1.0 ± 1.0) and averaged more than 3-fold the exposure of the radiologist (ratio, 3.2; 95% CI, 1.8-4.5). Multiple linear regression analysis showed that the exposure of the anesthesiologist was correlated with the number of pharmacologic interventions performed by the anesthesiologist and the total exposure of the radiologist.
CONCLUSIONS: Current guidelines for occupational radiation exposure to the eye are undergoing review and are likely to be lowered below the current 100-150 mSv/yr limit. Anesthesiologists who spend significant time in neurointerventional radiology suites may have ocular radiation exposure approaching that of a radiologist. To ensure parity with safety standards adopted by radiologists, these anesthesiologists should wear protective eyewear.

Entities:  

Mesh:

Year:  2011        PMID: 21285864     DOI: 10.1097/ALN.0b013e31820c2b81

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  11 in total

1.  Current attitudes of Turkish anesthesiologists to radiation exposure.

Authors:  Adnan Tüfek; Orhan Tokgöz; Ilker Öngüç Aycan; Feyzi Çelik; Abdurrahman Gümüş
Journal:  J Anesth       Date:  2013-05-08       Impact factor: 2.078

2.  Eye lens dosimetry in anesthesiology: a prospective study.

Authors:  Bart Vaes; Karel Van Keer; Lara Struelens; Werner Schoonjans; Ivo Nijs; Jan Vandevenne; Sven Van Poucke
Journal:  J Clin Monit Comput       Date:  2016-03-09       Impact factor: 2.502

3.  A survey on knowledge, attitude, and practices of workplace radiation safety amongst anaesthesiology personnel in northern Indian tertiary care institutes.

Authors:  Rudrashish Haldar; Rafat Shamim; Himel Mondal; Ashish Kumar Kannaujia; Prabhakar Mishra; Anil Agarwal
Journal:  Indian J Anaesth       Date:  2022-05-17

Review 4.  [Endovascular interventions in neuroradiology : New aspects].

Authors:  A Berlis
Journal:  Anaesthesist       Date:  2013-09       Impact factor: 1.041

Review 5.  Anesthesia for cardiac catheterization procedures.

Authors:  A Hamid
Journal:  Heart Lung Vessel       Date:  2014

6.  The radiation environment of anaesthesiologists in the endoscopic retrograde cholangiopancreatography room.

Authors:  Bora Lee; Min-Soo Kim; Darhae Eum; Kyeong Tae Min
Journal:  Sci Rep       Date:  2019-06-24       Impact factor: 4.379

Review 7.  Non-Operating Room Anesthesia: Patient Selection and Special Considerations.

Authors:  Timothy Wong; Paige L Georgiadis; Richard D Urman; Mitchell H Tsai
Journal:  Local Reg Anesth       Date:  2020-01-08

8.  Radiation Exposure during Neurointerventional Procedures in Modern Biplane Angiographic Systems: A Single-Site Experience.

Authors:  Ameer E Hassan; Sophie Amelot
Journal:  Interv Neurol       Date:  2017-02-11

9.  Interventional Radiologists: a Necessary Evaluation of Technical, Protective and Technological Operation.

Authors:  Sayed Ali Rahimi; Maryam Pourkaveh
Journal:  Acta Inform Med       Date:  2016-02-02

10.  Comparison of transesophageal and transthoracic echocardiography under moderate sedation for guiding transcatheter aortic valve replacement.

Authors:  Ines Sherifi; Alaa Mabrouk Salem Omar; Mithun Varghese; Menachem Weiner; Ani Anyanwu; Jason C Kovacic; Samin Sharma; Annapoorna Kini; Partho P Sengupta
Journal:  Echo Res Pract       Date:  2018-05-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.