Primary cup fixation with different designs of trapeziometacarpal total joint trapezium components: a radiostereometric analysis in a pig bone model.

T B Hansen1, M Meier, M C M Møller, K Larsen, M Stilling.   

Abstract

High loosening rates of the trapezium components of trapeziometacarpal total joint prostheses have been reported. The purpose of this study was to compare the primary press fit fixation of two different, uncemented cup designs (MOTEC and Elektra) with the primary fixation of a cemented polyethylene cup (DLC) in a pig bone model. We did a push out test to measure the maximal load strength of the implants and a low-pressure cyclic loading test combined with radiostereometric analysis to measure the micromotion of the implants. There was no significant difference in fixation strength between the two uncemented screw cup designs or between the two uncemented screw cups and the cemented polyethylene cup. However, we found that threading of the bone before insertion of the Electra screw cup weakens the primary fixation strength of the implant. The results indicate that focus should be on the insertion technique as well as on the cup design of uncemented trapezium cup implants. Further studies of trapezium implant migration in a clinical setting are needed.

Entities:  

Mesh:

Year:  2011        PMID: 21282215     DOI: 10.1177/1753193410394526

Source DB:  PubMed          Journal:  J Hand Surg Eur Vol        ISSN: 0266-7681


  4 in total

1.  Outcome of uncemented trapeziometacarpal prosthesis for treatment of thumb carpometacarpal joint arthritis.

Authors:  Manish Chug; Nicole Williams; David Benn; Stephen Brindley
Journal:  Indian J Orthop       Date:  2014-07       Impact factor: 1.251

Review 2.  The accuracy and precision of radiostereometric analysis in upper limb arthroplasty.

Authors:  Bart Ten Brinke; Annechien Beumer; Koen L M Koenraadt; Denise Eygendaal; Gerald A Kraan; Nina M C Mathijssen
Journal:  Acta Orthop       Date:  2017-03-02       Impact factor: 3.717

3.  Equally good fixation of cemented and uncemented cups in total trapeziometacarpal joint prostheses. A randomized clinical RSA study with 2-year follow-up.

Authors:  Torben Baek Hansen; Maiken Stilling
Journal:  Acta Orthop       Date:  2013-01-23       Impact factor: 3.717

4.  Trapezium Bone Density-A Comparison of Measurements by DXA and CT.

Authors:  Sebastian Breddam Mosegaard; Kamille Breddam Mosegaard; Nadia Bouteldja; Torben Bæk Hansen; Maiken Stilling
Journal:  J Funct Biomater       Date:  2018-01-18
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.