In vitro fluid dynamics of the Ahmed glaucoma valve modified with expanded polytetrafluoroethylene.

Francis Char DeCroos1, Yuji Kondo, Daniel Mordes, Maria Regina Lee, Sameer Ahmad, Sanjay Asrani, R Rand Allingham, Kevin C Olbrich, Bruce Klitzman.   

Abstract

PURPOSE: Long-term intraocular pressure reduction by glaucoma drainage devices (GDDs) is often limited by the fibrotic capsule that forms around them. Prior work demonstrates that modifying a GDD with a porous membrane promotes a vascularized and more permeable capsule. This work examines the in vitro fluid dynamics of the Ahmed valve after enclosing the outflow tract with a porous membrane of expanded polytetrafluoroethylene (ePTFE).
MATERIALS AND METHODS: The control and modified Ahmed implants (termed porous retrofitted implant with modified enclosure or PRIME-Ahmed) were submerged in saline and gelatin and perfused in a system that monitored flow (Q) and pressure (P). Flow rates of 1-50 μl/min were applied and steady state pressure recorded. Resistance was calculated by dividing pressure by flow.
RESULTS: Modifying the Ahmed valve implant outflow with expanded ePTFE increased pressure and resistance. Pressure at a flow of 2 μl/min was increased in the PRIME-Ahmed (11.6 ± 1.5 mm Hg) relative to the control implant (6.5 ± 1.2 mm Hg). Resistance at a flow of 2 μl/min was increased in the PRIME-Ahmed (5.8 ± 0.8 mm Hg/μl/min) when compared to the control implant (3.2 ± 0.6 mm Hg/μl/min).
CONCLUSIONS: Modifying the outflow tract of the Ahmed valve with a porous membrane adds resistance that decreases with increasing flow. The Ahmed valve implant behaves as a variable resistor. It is partially open at low pressures and provides reduced resistance at physiologic flow rates.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21281065     DOI: 10.3109/02713683.2010.512115

Source DB:  PubMed          Journal:  Curr Eye Res        ISSN: 0271-3683            Impact factor:   2.424


  2 in total

1.  Comparison of Silicone- and Porous-Plate Ahmed Glaucoma Valves.

Authors:  Tina M Roa; Peter A Netland; Vital P Costa; Steven R Sarkisian; Lama A Al-Aswad; Marlene R Moster; Iqbal I K Ahmed
Journal:  Med Devices (Auckl)       Date:  2020-07-16

Review 2.  Ahmed glaucoma valve implant: surgical technique and complications.

Authors:  Ivano Riva; Gloria Roberti; Francesco Oddone; Anastasios Gp Konstas; Luciano Quaranta
Journal:  Clin Ophthalmol       Date:  2017-02-17
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.