Safety and efficacy of electrospun polycarbonate-urethane vascular graft for early hemodialysis access: first clinical results in man.

Serosha M Wijeyaratne1, Lushanthi Kannangara.   

Abstract

OBJECTIVES: The purpose of this study was to assess the safety and efficacy of an electrospun multilayered, self-sealing polycarbonate-urethane graft for early hemodialysis access in patients.
METHOD: Seventeen eligible consenting patients had a polycarbonate-urethane graft (AVflo™) implanted and followed up prospectively for 12 months or to the end of secondary patency. Performance measures included graft patency, complications, time to first cannulation, and hemostasis times after needle withdrawal.
RESULTS: All patients were of Asian origin (mean age 57 years, range 29-78). Diabetes mellitus was the most common cause of renal failure (52.9%). There were no systemic or local reactions to the graft. Five patients (29.4%) died due to medical complications unrelated to the device. There was 1 pseudoaneurysm, 3 infected grafts that subsequently thrombosed, and 1 primary thrombosis associated with thrombophilia. One venous stenosis needed balloon angioplasty. Primary and secondary patency rates at 6 months were 72.7% and 81.8%, and at 12 months, 54.5% and 72.7%, respectively. Postimplantation vascular access needs were met entirely by the graft in every instance and prevented the need for venous catheters. Fifty-six percent were accessed within 8 days, the earliest being 48 hours. Finally, all arterial punctures and 98% of venous punctures had sealed in less than 5 minutes, with two thirds sealing off within 3 minutes of needle withdrawal.
CONCLUSION: The electrospun polycarbonate-urethane graft is safe in humans, permits early access obviating the need for venous catheters, and has equivalent patency to other prosthetic grafts at 1 year.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21279948     DOI: 10.5301/jva.2011.6278

Source DB:  PubMed          Journal:  J Vasc Access        ISSN: 1129-7298            Impact factor:   2.283


  4 in total

1.  Immediate use of an arteriovenous prosthetic graft for life-saving dialysis in a child.

Authors:  Chiara Grimaldi; Alessandro Crocoli; Lara De Galasso; Stefano Picca; Gian Luigi Natali; Jean De Ville De Goyet
Journal:  Pediatr Nephrol       Date:  2012-07-10       Impact factor: 3.714

Review 2.  In pursuit of functional electrospun materials for clinical applications in humans.

Authors:  Ryan J Stoddard; Arielle L Steger; Anna K Blakney; Kim A Woodrow
Journal:  Ther Deliv       Date:  2016-06-02

3.  An immediate access dialysis graft designed to prevent needle-related complications: Results from the initial pre-clinical studies.

Authors:  Shawn M Gage; Michael Lawson; Craig Nichols; Dalton Sycks; Roberto J Manson; Joseph A Knight
Journal:  J Vasc Access       Date:  2019-09-16       Impact factor: 2.283

4.  Computational study on the haemodynamic and mechanical performance of electrospun polyurethane dialysis grafts.

Authors:  Sjeng Quicken; Yeshi de Bruin; Barend Mees; Jan Tordoir; Tammo Delhaas; Wouter Huberts
Journal:  Biomech Model Mechanobiol       Date:  2019-11-02
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.