Herman Groepenhoff1, Caroline B Terwee, Patrick Mc Jak, Anton Vonk-Noordegraaf.
Abstract
BACKGROUND: To assess a pulmonary function change over time the mass flow sensor and the pneumotachograph are widely used in commercially available instruments. However, the smallest detectable change for both devices has never been compared. Therefore, the aim of this study is to determine the smallest detectable change in vital capacity (VC) and single-breath diffusion parameters measured by mass flow sensor and or pneumotachograph.Entities:
Year: 2011 PMID: 21276230 PMCID: PMC3236536 DOI: 10.1186/1756-0500-4-23
Source DB: PubMed Journal: BMC Res Notes ISSN: 1756-0500
Pulmonary function results on three different pulmonary function instruments.
| MF | PT |
| |
|---|---|---|---|
| 3 | 5 | ||
| 8 | 20 | ||
| 5.0 (0.7) | 4.7 (1.2) | ||
| 9.9 (2.3) | 9.0 (2.4) | ||
| 6.4 (1.0) | 5.8 (1.4) | ||
| 1.5 (0.3) | 1.6 (0.2) |
Results are expressed as mean and standard deviation. Mass flow sensor (MF); pneumotachograph (PT); VC: vital capacity; DLCO: transfer factor for carbon monoxide; VA: alveolar volume.
Figure 1Coefficient of variation (CV, left) and standard error of measurement (SEM, right) of vital capacity (VC), transfer factor for carbon monoxide (DLCO), alveolar volume (VA) and DLCO/VA. Each dot represents one hospital. MF is mass flow sensor; PT is pneumotachograph.
Figure 2Total error variance (= between apparatus, between occasion and random variation) for vital capacity (VC), transfer factor for carbon monoxide (DLCO), alveolar volume and DLCO/VA measured on three apparatuses. * p < 0.05 for random variation between mass flow sensor (MF) and pneumotachograph (PT) measurements.
Smallest detectable change measured by mass flow sensor (MF) or pneumotachograph(PT).
| MF | PT |
| |
|---|---|---|---|
| 0.53 (0.46 - 0.65) | 0.25 (0.17 - 0.36) | 0.04 | |
| 1.53 (1.26 - 1.70) | 1.18 (0.84 -1.39) | 0.07 | |
| 0.66 (0.53 -0.82) | 0.43 (0.34 - 0.53) | 0.04 | |
| 0.22 (0.19 - 0.28) | 0.19 (0.14 - 0.22) | 0.79 |
Results are expressed as mean and range. Mass flow sensor (MF); pneumotachograph (PT); VC: vital capacity; DLCO: transfer factor for carbon monoxide; VA: alveolar volume