| Literature DB >> 21276022 |
Claude S Elayi1, Matthew G Whitbeck, Gustavo Morales, John C Gurley.
Abstract
While modern implant tools have contributed greatly to the success of cardiac resynchronization therapy, technical challenges remain. A common problem is the inability to advance left ventricular pacing leads into branch veins that are tortuous or arise at steep angles. In these cases, advancement of the lead causes it to buckle and prolapse into the coronary sinus or great cardiac vein. Lead prolapsed can be avoided by employing a balloon to temporarily obstruct the coronary sinus or great cardiac vein just upstream from the branch vein. The balloon redirects the force of advancement laterally into the branch vein, facilitating delivery. ©2011, The Authors. Journal compilation ©2011 Wiley Periodicals, Inc.Mesh:
Year: 2011 PMID: 21276022 DOI: 10.1111/j.1540-8159.2010.03019.x
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976