Literature DB >> 21276022

Balloon-facilitated delivery of a left ventricular pacing lead.

Claude S Elayi1, Matthew G Whitbeck, Gustavo Morales, John C Gurley.   

Abstract

While modern implant tools have contributed greatly to the success of cardiac resynchronization therapy, technical challenges remain. A common problem is the inability to advance left ventricular pacing leads into branch veins that are tortuous or arise at steep angles. In these cases, advancement of the lead causes it to buckle and prolapse into the coronary sinus or great cardiac vein. Lead prolapsed can be avoided by employing a balloon to temporarily obstruct the coronary sinus or great cardiac vein just upstream from the branch vein. The balloon redirects the force of advancement laterally into the branch vein, facilitating delivery. ©2011, The Authors. Journal compilation ©2011 Wiley Periodicals, Inc.

Mesh:

Year:  2011        PMID: 21276022     DOI: 10.1111/j.1540-8159.2010.03019.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  A novel way to facilitate left ventricular lead implantation: Jailed catheter technique.

Authors:  Serkan Cay; Ozcan Ozeke; Firat Ozcan; Dursun Aras; Serkan Topaloglu
Journal:  J Arrhythm       Date:  2018-01-12
  1 in total