Aliasing artifacts with the BLADE technique: causes and effective suppression.

Shinya Kojima1, Satoru Morita, Eiko Ueno, Masami Hirata, Hiroyuki Shinohara, Akiyoshi Komori.   

Abstract

PURPOSE: To elucidate the causes of aliasing artifacts with the BLADE technique and clarify the effective suppression methods.
MATERIALS AND METHODS: We separately observed the aliasing artifacts of BLADE from features inside and then outside the defined field-of-view (FOV) using phantom studies. The effectiveness of suppressing them with phase oversampling (POS) and presaturation pulses (SAT) was evaluated. Finally, our observations were confirmed for a healthy volunteer.
RESULTS: Characteristic aliasing artifacts were observed from both inside and outside the FOV. Those from inside the FOV were sufficiently suppressed by using a POS of 25%, considering the acquisition time prolongation. Those from outside the FOV were nearly suppressed using SAT outside the FOV without selecting needless receiver coils. Aliasing artifacts on the coronal images of the head and neck with a healthy volunteer were completely suppressed by a combination of using a POS of 25% and using SAT on all four sides outside the FOV.
CONCLUSION: The characteristic aliasing artifacts of BLADE are caused from both inside and outside the defined FOV. They can be effectively suppressed by a combination of using a POS of 25% and using SAT on all four sides outside the FOV.
Copyright © 2011 Wiley-Liss, Inc.

Mesh:

Year:  2011        PMID: 21274986     DOI: 10.1002/jmri.22465

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  1 in total

1.  Robustness of a Combined Modified Dixon and PROPELLER Sequence with Two Interleaved Echoes in Clinical Head and Neck MRI.

Authors:  Yutaka Shigenaga; Daisuke Takenaka; Tomohisa Hashimoto; Takayuki Ishida
Journal:  Magn Reson Med Sci       Date:  2020-04-07       Impact factor: 2.471

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.