Reliability of bidirectional and variable-opening equipment for the measurement of pelvic floor muscle strength.

Fabiana Roberta Nunes1, Carla Campos Martins, Elaine Caldeira de Oliveira Guirro, Rinaldo Roberto Jesus Guirro.   

Abstract

OBJECTIVE: In evaluating pelvic floor muscles, it is important to use reliable and accurate methods. Therefore the objective of this study was to verify the reliability of bidirectional and variable-opening equipment designed to measure anteroposterior and left-right strength of pelvic floor muscles.
DESIGN: Test-retest to assess reliability.
SETTING: Academic institution, primary level of clinical care. PARTICIPANTS: Seventeen nulliparous women between 20 and 33 years of age participated in the procedure during 3 consecutive weeks.
METHODS: A dynamometer was used to evaluate pelvic floor muscle strength. The reliability was tested in 3 sessions, with a 7-day interval between, excluding premenstrual and menstrual periods. On each day of data collection, 3 maximum voluntary contractions of pelvic floor muscles were measured. The abdominal and gluteus muscles were evaluated concomitantly to the assessment in an attempt to isolate the pelvic floor muscle contractions. MAIN OUTCOME MEASURES: To evaluate the test-retest reliability with the intraclass correlation coefficient and the SEM.
RESULTS: Intraclass correlation of pelvic floor muscle force values ranged from good to excellent. The SEM values for anteroposterior strength were 1.96 N and 1.86 N for left and right, respectively.
CONCLUSIONS: Test-retest values demonstrated that the equipment we assessed to measure the anteroposterior and left-right force generated reliable pelvic floor muscle strength measurements. Copyright Â
© 2011 American Academy of Physical Medicine and Rehabilitation. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21257129     DOI: 10.1016/j.pmrj.2010.10.017

Source DB:  PubMed          Journal:  PM R        ISSN: 1934-1482            Impact factor:   2.298


  3 in total

1.  Subsequent Use of a Pressure Sensor to Record Intra-Abdominal Pressure After Maximum Vaginal Closure Force in a Clinical Trial.

Authors:  Stefan Niederauer; Brian Cottle; Xiaoming Sheng; James Ashton-Miller; John Delancey; Robert Hitchcock
Journal:  IEEE J Transl Eng Health Med       Date:  2019-11-15       Impact factor: 3.316

2.  Validity and reliability of an instrumented speculum designed to minimize the effect of intra-abdominal pressure on the measurement of pelvic floor muscle strength.

Authors:  James A Ashton-Miller; Ruth Zielinski; John O L DeLancey; Janis M Miller
Journal:  Clin Biomech (Bristol, Avon)       Date:  2014-10-02       Impact factor: 2.063

3.  Intra and inter-rater reliability study of pelvic floor muscle dynamometric measurements.

Authors:  Natalia M Martinho; Joseane Marques; Valéria R Silva; Silvia L A Silva; Leonardo C Carvalho; Simone Botelho
Journal:  Braz J Phys Ther       Date:  2015-04-27       Impact factor: 3.377

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.