Literature DB >> 21256318

Superior chest drainage with an active tube clearance system: evaluation of a downsized chest tube.

Yoko Arakawa1, Akira Shiose, Tohru Takaseya, Hideyuki Fumoto, Hyun-Il Kim, Edward M Boyle, A Marc Gillinov, Kiyotaka Fukamachi.   

Abstract

PURPOSE: We developed a small-diameter (20-French [Fr]) chest drainage system that incorporates an active tube clearance (ATC) system, and we evaluated its efficacy in an acute hemothorax model. DESCRIPTION: The ATC system includes an internal guidewire that can be advanced into the lumen of the chest tube to keep the tube from clogging. In six pigs, a 20-Fr ATC tube was placed on one side and a 32-Fr standard tube on the other, and 120 mL of fresh blood was infused into each pleural space every 15 minutes for a total of 840 mL. EVALUATION: The amount of drainage for 2 hours with a 20-Fr ATC tube was significantly greater than that with the 32-Fr standard tube (525 ± 179 mL vs 183 ± 85 mL; p = 0.0032).
CONCLUSIONS: This is the first time a smaller diameter tube has been shown to have better drainage in the setting of acute bleeding compared with larger diameter tubes. Copyright Â
© 2011 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21256318     DOI: 10.1016/j.athoracsur.2010.10.018

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  4 in total

1.  [Evaluation of the performance of a minimally invasive thoracic drainage tube in a rabbit model of hemothorax].

Authors:  Jia-Qing Zhang; Rui-Hong Ju; Kun-Tang Chen; Bao-Qin Ruan; Ting-Ting Xing
Journal:  Nan Fang Yi Ke Da Xue Xue Bao       Date:  2018-06-20

2.  Active Clearance of Chest Tubes Reduces Re-Exploration for Bleeding After Ventricular Assist Device Implantation.

Authors:  Simon Maltais; Mary E Davis; Nicholas A Haglund; Louis Perrault; Sudhir S Kushwaha; John M Stulak; Edward M Boyle
Journal:  ASAIO J       Date:  2016 Nov/Dec       Impact factor: 2.872

3.  Active clearance of chest tubes is associated with reduced postoperative complications and costs after cardiac surgery: a propensity matched analysis.

Authors:  Yvon Baribeau; Benjamin Westbrook; Yanick Baribeau; Simon Maltais; Edward M Boyle; Louis P Perrault
Journal:  J Cardiothorac Surg       Date:  2019-11-08       Impact factor: 1.637

4.  Continuous postoperative pericardial flushing reduces postoperative bleeding after coronary artery bypass grafting: A randomized trial.

Authors:  Eva C Diephuis; Corianne A de Borgie; A Zwinderman; Jacobus A Winkelman; Wim-Jan P van Boven; José P S Henriques; Susanne Eberl; Nicole P Juffermans; Marcus J Schultz; Robert J M Klautz; David R Koolbergen
Journal:  EClinicalMedicine       Date:  2020-12-23