Literature DB >> 21255511

Pilot study of a new device to titrate oxygen flow in hypoxic patients on long-term oxygen therapy.

Serena Cirio1, Stefano Nava.   

Abstract

BACKGROUND: The O(2) Flow Regulator (Dima, Bologna, Italy) is a new automated oxygen regulator that titrates the oxygen flow based on a pulse-oximetry signal to maintain a target S(pO(2)). We tested the device's safety and efficacy.
METHODS: We enrolled 18 subjects with chronic lung disease, exercise-induced desaturation, and on long-term oxygen therapy, in a randomized crossover study with 2 constant-work-load 15-min cycling exercise tests, starting with the patient's previously prescribed usual oxygen flow. In one test the oxygen flow was titrated manually by the respiratory therapist, and in the other test the oxygen flow was titrated by the O(2) Flow Regulator, to maintain an S(pO(2)) of 94%. We measured S(pO(2)) throughout each test, the time spent by the respiratory therapist to set the device or to manually regulate the oxygen flow, and the total number of respiratory-therapist titration interventions during the trial.
RESULTS: There were no differences in symptoms or heart rate between the exercise tests. Compared to the respiratory-therapist-controlled tests, during the O(2) Flow Regulator tests S(pO(2)) was significantly higher (95 ± 2% vs 93 ± 3%, P = .04), significantly less time was spent below the target S(pO(2)) (171 ± 187 s vs 340 ± 220 s, P < .001), and the O(2) Flow Regulator tests required significantly less respiratory therapist time (5.6 ± 3.7 min vs 2.0 ± 0.1 min, P = .005).
CONCLUSIONS: The O(2) Flow Regulator may be a safe and effective alternative to manual oxygen titration during exercise in hypoxic patients. It provided stable S(pO(2)) and avoided desaturations in our subjects.

Entities:  

Mesh:

Year:  2011        PMID: 21255511     DOI: 10.4187/respcare.00983

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  9 in total

Review 1.  Optimal oxygen titration in patients with chronic obstructive pulmonary disease: a role for automated oxygen delivery?

Authors:  François Lellouche; Jed Lipes; Erwan L'Her
Journal:  Can Respir J       Date:  2013 Jul-Aug       Impact factor: 2.409

2.  Automated oxygen titration and weaning with FreeO2 in patients with acute exacerbation of COPD: a pilot randomized trial.

Authors:  François Lellouche; Pierre-Alexandre Bouchard; Maude Roberge; Serge Simard; Erwan L'Her; François Maltais; Yves Lacasse
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2016-08-24

3.  Cost-effectiveness of FreeO2 in patients with chronic obstructive pulmonary disease hospitalised for acute exacerbations: analysis of a pilot study in Quebec.

Authors:  Thomas G Poder; Christian R C Kouakou; Pierre-Alexandre Bouchard; Véronique Tremblay; Sébastien Blais; François Maltais; François Lellouche
Journal:  BMJ Open       Date:  2018-01-23       Impact factor: 2.692

4.  Automated oxygen administration versus conventional oxygen therapy after major abdominal or thoracic surgery: study protocol for an international multicentre randomised controlled study.

Authors:  Erwan L'her; Samir Jaber; Daniel Verzilli; Christophe Jacob; Brigitte Huiban; Emmanuel Futier; Thomas Kerforne; Victoire Pateau; Pierre-Alexandre Bouchard; Maellen Gouillou; Emmanuel Nowak; François Lellouche
Journal:  BMJ Open       Date:  2019-01-17       Impact factor: 2.692

5.  Automated oxygen control with O2matic® during admission with exacerbation of COPD.

Authors:  Ejvind Frausing Hansen; Jens Dahlgaard Hove; Charlotte Sandau Bech; Jens-Ulrik Stæhr Jensen; Thomas Kallemose; Jørgen Vestbo
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2018-12-14

6.  Automated Oxygen Delivery in Hospitalized Patients with Acute Respiratory Failure: A Pilot Study.

Authors:  Foteini Malli; Stelios Boutlas; Nick Lioufas; Konstantinos I Gourgoulianis
Journal:  Can Respir J       Date:  2019-02-03       Impact factor: 2.409

7.  Feasibility and reliability of an automated controller of inspired oxygen concentration during mechanical ventilation.

Authors:  Kaouther Saihi; Jean-Christophe M Richard; Xavier Gonin; Thomas Krüger; Michel Dojat; Laurent Brochard
Journal:  Crit Care       Date:  2014-02-19       Impact factor: 9.097

8.  Portable Automated Oxygen Administration System for hypoxaemic patients.

Authors:  Khawla Alzoubi; Ziyad Alguraan; Omar M Ramahi
Journal:  Springerplus       Date:  2016-04-18

9.  Evaluation of a Novel Ear Pulse Oximeter: Towards Automated Oxygen Titration in Eyeglass Frames.

Authors:  Fabian Braun; Christophe Verjus; Josep Solà; Marcus Marienfeld; Manuela Funke-Chambour; Jens Krauss; Thomas Geiser; Sabina A Guler
Journal:  Sensors (Basel)       Date:  2020-06-10       Impact factor: 3.576

  9 in total