Literature DB >> 21242679

Evaluation of a new online hemodiafiltration mode with automated pressure control of convection.

U Teatini1, D Steckiph, G Romei Longhena.   

Abstract

BACKGROUND: Postdilution hemodiafiltration (HDF) still remains the gold standard for solute removal, but with high hemoglobin levels transmembrane pressure (TMP) may reach high levels, reducing filter performance. We compared three online postdilution HDF treatments without TMP alarms for convective volume and plasma changes following treatment.
METHODS: Twelve patients were enrolled in a trial with three online postdilution HDF treatments. In the volume-controlled mode (VOLc), we set the exchanged volume to obtain a filtration fraction close to 25% without TMP alarms. In the pressure-controlled mode (TMPc), we set the TMP at 100 mm Hg. In the ULTRAc mode (TMP scan combined with TMPc), the dialysis machine automatically selects the TMP with a scan. All treatments were performed with an AK200 ULTRA-S system.
RESULTS: Even with hemoglobin levels >12 g/dl, we found a 57% rise in ultrafiltration rate in TMPc versus VOLc and a 92% rise in ULTRAc versus VOLc. Phosphates and myoglobin levels were significantly affected by treatment type.
CONCLUSIONS: ULTRAc may be a useful tool to achieve an excellent purification performance without the constraints associated with the risk of hemoconcentration.
Copyright © 2011 S. Karger AG, Basel.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21242679     DOI: 10.1159/000321884

Source DB:  PubMed          Journal:  Blood Purif        ISSN: 0253-5068            Impact factor:   2.614


  2 in total

1.  Optimization of the convection volume in online post-dilution haemodiafiltration: practical and technical issues.

Authors:  Isabelle Chapdelaine; Camiel L M de Roij van Zuijdewijn; Ira M Mostovaya; Renée Lévesque; Andrew Davenport; Peter J Blankestijn; Christoph Wanner; Menso J Nubé; Muriel P C Grooteman; P J Blankestijn; A Davenport; C Basile; F Locatelli; F Maduell; S Mitra; C Ronco; R Shroff; J Tattersall; C Wanner
Journal:  Clin Kidney J       Date:  2015-02-16

2.  Achieving high convection volumes in postdilution online hemodiafiltration: a prospective multicenter study.

Authors:  Camiel L M de Roij van Zuijdewijn; Isabelle Chapdelaine; Menso J Nubé; Peter J Blankestijn; Michiel L Bots; Constantijn J A M Konings; Ton K Kremer Hovinga; Femke M Molenaar; Neelke C van der Weerd; Muriel P C Grooteman
Journal:  Clin Kidney J       Date:  2017-02-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.