In vitro analysis of resistance to cyclic load and preload distribution of two implant/abutment screwed connections.

Giovanna Murmura1, Donato Di Iorio, Angelo Raffaele Cicchetti, Bruna Sinjari, Sergio Caputi.   

Abstract

The aim of the present research is an in vitro evaluation of the preload distribution in screw-retained implant systems under cyclic load. Two implant systems with internal connection were tested: fifteen 4.5 × 10 mm implants with internal hexagon and fifteen 4.5 × 10 mm implants with internal octagon. Samples underwent cyclic load that was between 20 N and 200 N for 1 × 10(6) cycles. After mechanical tests, samples were sectioned along the long axis and analyzed under a scanning electron microscope. Five 4.5 × 10 mm implants with internal hexagon and five 4.5 × 10 mm implants with internal octagon were collected for photoelastic analysis. Each fixture was mounted in a wax-made parallelepiped measuring 20 mm × 20 mm × 10 mm. A mold was made for each wax parallelepiped/fixture assembly using a silicone-based impression material, and an epoxy resin was poured in each mold. After setting of the resin, 25° angled titanium abutments were screwed onto each replica; afterwards, assemblies underwent photoelastic analysis. After cyclic load, screw threads and heads were still in contact with internal fixture threads and abutment holes, respectively, suggesting that preload has not been lost during load. During load, SSO and Xsigñ implants behave in a different way. SSO samples revealed the presence of fringes radiating from the base of the abutment. Xsigñ implants showed the presence of fringes radiating from the threads of the retention screw. From the present in vitro research, it is possible to state that screw-retained abutment based on an internal octagonal connection is less likely to come loose after cyclic load.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21231867     DOI: 10.1563/AAID-JOI-D-10-00144

Source DB:  PubMed          Journal:  J Oral Implantol        ISSN: 0160-6972            Impact factor:   1.779


  4 in total

Review 1.  Preload loss of abutment screws after dynamic fatigue in single implant-supported restorations. A systematic review.

Authors:  Beatriz Pardal-Peláez; Javier Montero
Journal:  J Clin Exp Dent       Date:  2017-11-01

Review 2.  Systematic analysis of factors that cause loss of preload in dental implants.

Authors:  S Nithyapriya; A S Ramesh; A Kirubakaran; Jeevitha Mani; J Raghunathan
Journal:  J Indian Prosthodont Soc       Date:  2018 Jul-Sep

Review 3.  Review of the Mechanical Behavior of Different Implant-Abutment Connections.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Paula López-Jarana; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-11-23       Impact factor: 3.390

4.  Mechanical Pull-Out Test of a New Hybrid Fixture-Abutment Connection: An In Vitro Study.

Authors:  Gianmaria D'Addazio; Bruna Sinjari; Lorenzo Arcuri; Beatrice Femminella; Giovanna Murmura; Manlio Santilli; Sergio Caputi
Journal:  Materials (Basel)       Date:  2021-03-22       Impact factor: 3.623

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.