Evaluation of the Pentax-AWS(®) and the Macintosh laryngoscope in difficult intubation: a manikin study.

T Kariya1, G Inagawa, K Nakamura, J Fujimoto, Y Aoi, S Morita, T Goto.   

Abstract

BACKGROUND: The Pentax-AWS (AWS(®)), a new video laryngoscope, has been shown to be useful in cases of difficult intubation. We hypothesized that the AWS(®) would be more useful in the settings of a narrow upper airway than the Macintosh laryngoscope. We compared each device in simulated scenarios of representative difficulty of tracheal intubation using a manikin. The primary endpoint was the rate of successful intubation.
METHODS: With each device, 23 anesthesiologists performed tracheal intubation in a SimMan(®) manikin in the following scenarios: (1) normal airway, (2) tongue edema, (3) cervical spine rigidity, (4) pharyngeal obstruction, (5) jaw trismus, (6) tongue edema with pharyngeal obstruction. The intubation time and success rate were measured. Each participant was asked to rate the difficulty of intubation (1=very easy; 5=very difficult).
RESULTS: In the scenarios of tongue edema and tongue edema with pharyngeal obstruction, the AWS(®) yielded a higher success rate (100% vs. 34.8%; P<0.001, 65.2% vs. 21.7%; P=0.006), a shorter intubation time [14.6 (7.0) vs. 33.4 (13.0) s; P<0.001, 24.5 (12.0) vs. 37.6 (11.9); P=0.047; mean (standard deviation)], and a lower difficulty score [2 (1-4) vs. 5 (1-5); P<0.001, 4 (2-5) vs. 5 (3-5); P<0.001; median (range)], compared with the Macintosh laryngoscope.
CONCLUSION: The AWS(®) has an advantage over the Macintosh laryngoscope in simulated tongue edema and tongue edema with pharyngeal obstruction. Further studies in a clinical setting are necessary to confirm these findings.
© 2011 The Authors. Journal compilation © 2011 The Acta Anaesthesiologica Scandinavica Foundation.

Entities:  

Mesh:

Year:  2011        PMID: 21226865     DOI: 10.1111/j.1399-6576.2010.02363.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  2 in total

1.  Supraglottic airway device versus a channeled or non-channeled blade-type videolaryngoscope for accidental extubation in the prone position: A randomized crossover manikin study.

Authors:  Hiroyuki Oshika; Yukihide Koyama; Masataka Taguri; Koichi Maruyama; Go Hirabayashi; Shoko Merrit Yamada; Masashi Kohno; Tomio Andoh
Journal:  Medicine (Baltimore)       Date:  2018-06       Impact factor: 1.889

2.  Usefulness of video-laryngoscopy with the Airway Scope for intubation performance and learning: an experimental manikin controlled study.

Authors:  Pierre-Louis Declercq; Michael Bubenheim; Stéphanie Gelinotte; Kévin Guernon; Jean-Baptiste Michot; Vincent Royon; Dorothée Carpentier; Gaëtan Béduneau; Fabienne Tamion; Christophe Girault
Journal:  Ann Intensive Care       Date:  2016-08-26       Impact factor: 6.925

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.