Optimal continuous positive airway pressure level in korean patients with obstructive sleep apnea syndrome.

Ji Ho Choi1, Eun Joong Kim, Kang Woo Kim, June Choi, Soon Young Kwon, Heung Man Lee, Tae Hoon Kim, Sang Hag Lee, Chol Shin, Seung Hoon Lee.   

Abstract

OBJECTIVES: The aim of this study was to investigate optimal continuous positive airway pressure (CPAP) level, to examine the factors affecting optimal CPAP level, and to develop a predictive equation for optimal CPAP level in Korean patients with obstructive sleep apnea syndrome (OSAS).
METHODS: A total of 202 patients with OSAS who underwent successful manual titration for CPAP treatment were included in this study. Correlations between the optimal CPAP level and baseline data including anthropometric and polysomnographic variables were analyzed. A predictive equation for optimal CPAP level was developed based on anthropometric and polysomonographic data.
RESULTS: The mean optimal CPAP level in 202 patients with OSAS was 7.8±2.3 cm H(2)O. The mean optimal CPAP level in the mild, moderate, and severe OSAS groups was 6.0±1.3, 7.4±1.9, and 9.1±2.1 cm H(2)O, respectively. The apneahypopnea index (AHI) (r=0.595, P<0.001), arousal index (r=0.542, P<0.001), minimal SaO(2) (r=-0.502, P<0.001), body mass index (BMI) (r=0.494, P<0.001), neck circumference (r=0.265, P<0.001), and age (r=-0.164, P=0.019) were significantly correlated with optimal CPAP level. The best predictive equation according to stepwise multiple linear regression analysis was: Optimal CPAP level (cm H(2)O)=0.681+(0.205×BMI)+(0.040×AHI). Forty-two percent of the variance in the optimal CPAP level was explained by this equation (R(2)=0.42, P<0.001).
CONCLUSION: A predictive equation for optimal CPAP level in Korean patients with OSAS was developed using AHI and BMI, which can be easily measured during the diagnostic process.

Entities:  

Keywords:  Body mass index; Continuous positive airway pressure; Obstructive sleep apnea syndrome; Polysomnography

Year:  2010        PMID: 21217962      PMCID: PMC3010540          DOI: 10.3342/ceo.2010.3.4.207

Source DB:  PubMed          Journal:  Clin Exp Otorhinolaryngol        ISSN: 1976-8710            Impact factor:   3.372


INTRODUCTION

Obstructive sleep apnea syndrome (OSAS) is a common chronic disease that requires long-term treatment and follow-up (1). The estimated prevalence of OSAS in middle-aged Koreans is 4.5% in males and 3.2% in females (2). Treatment for OSAS consists of medical, behavioral and surgical therapy (1, 3). Medical therapy includes continuous positive airway pressure (CPAP) and oral appliances, and behavioral therapy consists of weight loss and positional therapy (1, 4-8). In 1981, Sullivan et al. (4) first introduced nasal CPAP, working as a pneumatic splint to prevent upper airway collapse, for OSAS treatment. CPAP is routinely recommended for treatment in patients with moderate to severe OSAS and optionally indicated for the treatment of patients with mild OSAS (9). After CPAP treatment, benefits in patients with OSAS include improvement of self-reported daytime sleepiness and quality of life (10, 11). An optimal CPAP level is important. A lower than optimal CPAP level may result in insufficient treatment effects and/or unintentional mask removal, while a higher than optimal CPAP level may induce pressure intolerance and/or mouth dryness (12). Three titration methods are used to establish an optimal level: full-night, split-night, and titration using auto-titrating positive airway pressure (APAP). Of these, full-night, attended titration with polysomnography (PSG) is the gold standard to determine an optimal CPAP level (1). Although full-night manual titration in a sleep laboratory is the preferred method, APAP titration may be performed to identify optimal CPAP level for patients with moderate to severe OSAS without congestive heart failure, chronic obstructive pulmonary disease, or hypoventilation syndrome (13). Various predictive equations for optimal CPAP level have been developed in several countries and for populations of different geographical origins (14-20). The optimal CPAP level determined by a predictive equation may improve the success rate of manual titration and increase the convenience of treatments such as patient self-titration of CPAP or APAP (21-23). Therefore, the objectives of the present study were: 1) to investigate mean optimal CPAP level according to the severity of OSAS; 2) to examine the factors which affect optimal CPAP level; and 3) ultimately to develop predictive equation for optimal CPAP level using variables which can be easily measured during the diagnostic process in Korean patients with OSAS.

MATERIALS AND METHODS

Subjects

Two hundred two Korean patients diagnosed with OSAS using the apnea-hypopnea index (AHI) in standard PSG ≥5 with signs and symptoms suspicious for OSAS, and who underwent a successful manual titration for CPAP treatment between April 2002 and December 2004 were included in this study. The subjects comprised 182 males and 20 females with mild (n=54), moderate (n=49), and severe (n=99) OSAS (Table 1). Mean age and body mass index (BMI, weight in kg divided by height in m2) was 44.8±9.3 years and 27.6±3.4 kg/m2, respectively. Severity of OSAS was classified according to the degree of AHI. Mild, moderate, and severe OSAS for adults were defined as 5≤AHI<15, 15≤AHI<30, AHI≥30, respectively (24). Neck circumference (cm) was measured at the level just below the most prominent portion of the thyroid cartilage (Adam's apple). The study was reviewed and approved by the Institutional Review Board at Ansan Hospital, Korea University.
Table 1

Anthropometric and polysomonographic variables according to the severity of OSAS

Data are presented as mean±SD.

OSAS: obstructive sleep apnea syndrome; TST: total sleep time; REM: rapid eye movement; AHI: apnea-hypopnea index.

PSG and CPAP titration

All subjects underwent an overnight, in-laboratory attended PSG with an Alice 4 computerized polysomnographic system (Respironics, Atlanta, GA, USA). Sixteen channels were simultaneously used to document the following parameters: four channel electroencephalogram, electrooculogram, submental and leg electromyogram, electrocardiogram, airflow at the nose and mouth (thermistor, nasal pressure transducer), chest and abdominal respiratory movement, oxygen saturation (pulse oximetry), snoring microphone, and body position sensor. A sleep technician observed the behavior of the subjects and confirmed their sleep positions using an infrared camera inside the room. To determine optimal pressure level for CPAP treatment, fullnight, manual titration with the same montage PSG was performed. The optimal pressure level was defined as the lowest effective pressure that controlled most respiratory disturbances including apnea, hypopnea, and snoring in all body positions and in all stages, especially in the supine position during rapid eye movement (REM) sleep. All the sleep studies were manually interpreted by a sleep technician according to the standard sleep stage and arousal criteria (25, 26). Apnea was defined as absence of airflow for a period lasting at least 10 seconds and hypopnea was defined as at least a 30% reduction in airflow associated with decrease in oxygen saturation of 4% or more. The apnea index was defined as the number of apneic episodes per hour of total sleep time, and the AHI was defined as the number of episodes of apnea and hypopnea per hour of total sleep time. Arousal index was defined as the number of arousals per hour of total sleep time.

Statistical analysis

All data are presented as the mean±standard deviation for continuous variables, and as frequencies (percent) for categorical variables. Pearson's correlation coefficients were used to analyze relationships between optimal CPAP level and baseline data including anthropometric and polysomnographic variables. Stepwise multiple linear regression analysis was used to select independent predictive variables and to develop a predictive equation for optimal CPAP level. Statistical analysis was performed using SPSS ver. 10.0 (SPSS Inc., Chicago, IL, USA), and P<0.05 was considered statistically significant.

RESULTS

The mean optimal pressure level for CPAP treatment according to the severity of OSAS is shown in Table 2. The mean optimal pressure level in the 202 OSAS patients was 7.8±2.3 cm H2O. The mean optimal pressure levels in the mild, moderate, and severe OSAS groups was 6.0±1.3, 7.4±1.9, and 9.1±2.1 cm H2O, respectively.
Table 2

Mean optimal CPAP level according to the severity of OSAS

Data are presented as mean±SD.

CPAP: continuous positive airway pressure; OSAS: obstructive sleep apnea syndrome.

Correlations between optimal pressure level for CPAP treatment and baseline variables including demographic and polysomonographic data in the 202 patients are presented in Table 3. AHI (r=0.595, P<0.001), arousal index (r=0.542, P<0.001), minimal SaO2 (r=-0.502, P<0.001), BMI (r=0.494, P<0.001), neck circumference (r=0.265, P<0.001), and age (r=-0.164, P=0.019) were significantly correlated with optimal pressure level.
Table 3

Relationship between optimal CPAP level and baseline variables in 202 patients with OSAS

Data are presented as mean±SD.

CPAP: continuous positive airway pressure; OSAS: obstructive sleep apnea syndrome; TST: total sleep time; AHI: apnea-hypopnea index; NS: no significance.

Of the anthropometric and polysomnographic variables, stepwise multiple linear regression analysis identified BMI and AHI as independent predictive variables associated with the optimal pressure level. The best predictive equation by stepwise regression analysis was: Optimal CPAP level (cm H2O)=0.681+(0.205×BMI)+(0.040×AHI) Forty-two percent of the variance in the optimal CPAP level was explained by this equation (R2=0.42, P<0.001).

DISCUSSION

The ultimate objective of this study, to develop a predictive equation for optimal CPAP level using easily measured variables during diagnostic evaluation in Korean OSAS patients, was accomplished. Lin et al. (17) constructed a predictive model for optimal CPAP level in 121 adult Taiwanese patients with OSAS and developed the following equation: Optimal CPAP level (cm H2O)=0.52+(0.174×BMI)+(0.042×AHI). Their predictive model is in close agreement with our equation, in that BMI and AHI were the two most important independent predictors of optimal CPAP level. Akahoshi et al. (20) produced a predictive equation derived from anthropometric, polysomnographic, and cephalometric data in 170 Japanese patients with OSAS. In the study, stepwise multiple regression analysis selected BMeH (the angle between a line from point to menton and a line from Me to the hyoid bone ), BMI, AHI, and mean SaO2 as independent predictors of effective pressure. The predictive equation was: Optimal CPAP level (cm H2O)=27.78+(0.041×BMeH)+(0.141×BMI)+(0.040×AHI)-(0.312×mean SaO2). Although optimal pressure was more accurately predicted by combining cephalometric data such as BMeH with BMI, AHI, and mean SaO2, cephalometric analysis may decrease the clinical usefulness of this equation. Rowley et al. (21) investigated the effect of predictive equations on the success rate of manual titration for CPAP treatment. They found that manual titration using predictive equation modestly increases the success rate of CPAP titration. The success rate of titration studies for which the AHI at the final level tested ≤50% of baseline and ≤10 events/hour increased from 50% to 68% (P<0.001). It is thought that the predictive equation worked by reducing complex steps and time-consuming efforts to determine effective pressure levels. Fitzpatrick et al. (22) compared results between in-laboratory manual titration and home self-titration of CPAP for OSAS in a randomized, singleblind, two-period crossover study. Patient self-titration of CPAP was performed based on the optimal CPAP level by the predictive equation. The authors demonstrated that home self-titration of CPAP using a predictive equation is as effective as in-laboratory manual titration during full-night PSG, with similar CPAP compliance, subjective, and objective outcomes. Predictive equations for optimal CPAP level may be required for the following objectives: 1) to improve the success rate of full-night, in-laboratory manual titration, because the starting pressure by predictive equation would be closer to the optimal CPAP level (e.g., simplifying the titration process as decreasing numerous pressure changes and time consumption to obtain an optimal pressure level); 2) to increase the convenience and effectiveness of home patient management (e.g., home self-titration of CPAP or APAP); 3) to be substituted for manual titration in several conditions in which manual titration in a sleep laboratory is not possible (e.g., immobility, safety, and critical illness); and 4) as an alternative to manual titration in cases in which titration using a predictive equation would be preferable to manual titration (e.g., high cost, time consumption, and long waiting list). The strength of the current study is that the study sample was large and included patients with mild OSAS. In two previous East Asian studies, the study samples were 121 patients with moderate to severe OSAS and 170 patients with an AHI of more than 20 (17, 20). CPAP is optionally indicated for the treatment of patients with mild OSAS. We believe that a better predictive equation may be developed for the large number of patients with OSAS, including those with mild OSAS. There were several limitations in the present study. This study did not include a validation or application of the presently developed equation to another group of Korean OSAS patients. Although cephalometric data may influence optimal pressure level, craniofacial abnormalities were not considered in the development of this predictive equation because the required cephalometric analysis may have diminished the clinical utility of this equation. In conclusion, a predictive equation for optimal CPAP level was developed using AHI and BMI, which can be easily measured during diagnostic work-up. We expect that this predictive equation for optimal CPAP level will be helpful for designing CPAP treatment in Korean patients with OSAS. CONFLICT OF INTEREST No potential conflict of interest relevant to this article was reported.
  25 in total

1.  Proposed supplements and amendments to 'A Manual of Standardized Terminology, Techniques and Scoring System for Sleep Stages of Human Subjects', the Rechtschaffen & Kales (1968) standard.

Authors:  T Hori; Y Sugita; E Koga; S Shirakawa; K Inoue; S Uchida; H Kuwahara; M Kousaka; T Kobayashi; Y Tsuji; M Terashima; K Fukuda; N Fukuda
Journal:  Psychiatry Clin Neurosci       Date:  2001-06       Impact factor: 5.188

2.  Predicting optimal continuous positive airway pressure in Japanese patients with obstructive sleep apnoea syndrome.

Authors:  Toshiki Akahoshi; Tsuneto Akashiba; Seiji Kawahara; Akihito Uematsu; Kenichi Nagaoka; Kouji Kiyofuji; Naoki Okamoto; Tomohiro Hattori; Noriaki Takahashi; Shu Hashimoto
Journal:  Respirology       Date:  2008-12-11       Impact factor: 6.424

Review 3.  Practice parameters for the indications for polysomnography and related procedures: an update for 2005.

Authors:  Clete A Kushida; Michael R Littner; Timothy Morgenthaler; Cathy A Alessi; Dennis Bailey; Jack Coleman; Leah Friedman; Max Hirshkowitz; Sheldon Kapen; Milton Kramer; Teofilo Lee-Chiong; Daniel L Loube; Judith Owens; Jeffrey P Pancer; Merrill Wise
Journal:  Sleep       Date:  2005-04       Impact factor: 5.849

4.  Comparison of therapeutic and subtherapeutic nasal continuous positive airway pressure for obstructive sleep apnoea: a randomised prospective parallel trial.

Authors:  C Jenkinson; R J Davies; R Mullins; J R Stradling
Journal:  Lancet       Date:  1999-06-19       Impact factor: 79.321

Review 5.  Prediction of continuous positive airway pressure in obstructive sleep apnea.

Authors:  José S Loredo; Charles Berry; Richard A Nelesen; Joel E Dimsdale
Journal:  Sleep Breath       Date:  2007-03       Impact factor: 2.816

6.  Reversal of obstructive sleep apnoea by continuous positive airway pressure applied through the nares.

Authors:  C E Sullivan; F G Issa; M Berthon-Jones; L Eves
Journal:  Lancet       Date:  1981-04-18       Impact factor: 79.321

7.  Relative accuracy of algorithm-based prescription of nasal CPAP in OSA.

Authors:  J R Stradling; M Hardinge; J Paxton; D M Smith
Journal:  Respir Med       Date:  2004-02       Impact factor: 3.415

8.  Practice parameters for the treatment of snoring and Obstructive Sleep Apnea with oral appliances: an update for 2005.

Authors:  Clete A Kushida; Timothy I Morgenthaler; Michael R Littner; Cathy A Alessi; Dennis Bailey; Jack Coleman; Leah Friedman; Max Hirshkowitz; Sheldon Kapen; Milton Kramer; Teofilo Lee-Chiong; Judith Owens; Jeffrey P Pancer
Journal:  Sleep       Date:  2006-02       Impact factor: 5.849

9.  Practice parameters for the use of continuous and bilevel positive airway pressure devices to treat adult patients with sleep-related breathing disorders.

Authors:  Clete A Kushida; Michael R Littner; Max Hirshkowitz; Timothy I Morgenthaler; Cathy A Alessi; Dennis Bailey; Brian Boehlecke; Terry M Brown; Jack Coleman; Leah Friedman; Sheldon Kapen; Vishesh K Kapur; Milton Kramer; Teofilo Lee-Chiong; Judith Owens; Jeffrey P Pancer; Todd J Swick; Merrill S Wise
Journal:  Sleep       Date:  2006-03       Impact factor: 5.849

10.  Practice parameters for the use of autotitrating continuous positive airway pressure devices for titrating pressures and treating adult patients with obstructive sleep apnea syndrome: an update for 2007. An American Academy of Sleep Medicine report.

Authors:  Timothy I Morgenthaler; R Nisha Aurora; Terry Brown; Rochelle Zak; Cathy Alessi; Brian Boehlecke; Andrew L Chesson; Leah Friedman; Vishesh Kapur; Rama Maganti; Judith Owens; Jeffrey Pancer; Todd J Swick
Journal:  Sleep       Date:  2008-01       Impact factor: 5.849

View more
  7 in total

1.  A predictive model for optimal continuous positive airway pressure in the treatment of pure moderate to severe obstructive sleep apnea in China.

Authors:  Le Wang; Xing Chen; Dong-Hui Wei; Mao-Li Liang; Yan Wang; Bao-Yuan Chen; Jing Zhang; Jie Cao
Journal:  BMC Pulm Med       Date:  2022-06-16       Impact factor: 3.320

2.  Upper airway anatomical balance contributes to optimal continuous positive airway pressure for Japanese patients with obstructive sleep apnea syndrome.

Authors:  Eiki Ito; Satoru Tsuiki; Kazuyoshi Namba; Yuji Takise; Yuichi Inoue
Journal:  J Clin Sleep Med       Date:  2014-02-15       Impact factor: 4.062

3.  Titration studies overestimate continuous positive airway pressure requirements in uncomplicated obstructive sleep apnea.

Authors:  Olabimpe S Fashanu; Rohit Budhiraja; Salma Batool-Anwar; Stuart F Quan
Journal:  J Clin Sleep Med       Date:  2021-09-01       Impact factor: 4.324

4.  Are there sex-related differences in therapeutic CPAP levels in adults undergoing in-laboratory titration?

Authors:  Ricardo L M Duarte; Flavio J Magalhães-da-Silveira; David Gozal
Journal:  J Clin Sleep Med       Date:  2021-09-01       Impact factor: 4.324

Review 5.  Mathematical Equations to Predict Positive Airway Pressures for Obstructive Sleep Apnea: A Systematic Review.

Authors:  Macario Camacho; Muhammad Riaz; Armin Tahoori; Victor Certal; Clete A Kushida
Journal:  Sleep Disord       Date:  2015-07-30

6.  Predictive equation for optimal continuous positive airway pressure in children with obstructive sleep apnoea.

Authors:  Joelle Chong; Ram Bajpai; Oon Hoe Teoh; Arun Pugalenthi; John C Allen; Zai Ru Cheng; Soh Gin Tan; Meilan Lim; Jasmine Tan; Anne Goh; Yi Hua Tan; Biju Thomas
Journal:  ERJ Open Res       Date:  2020-06-08

7.  Using craniofacial characteristics to predict optimum airway pressure in obstructive sleep apnea treatment.

Authors:  Thays Crosara Abrahão Cunha; Thais Moura Guimarães; Fernanda R Almeida; Fernanda L M Haddad; Luciana B M Godoy; Thulio M Cunha; Luciana O Silva; Sergio Tufik; Lia Bittencourt
Journal:  Braz J Otorhinolaryngol       Date:  2018-12-12
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.