| Literature DB >> 21217912 |
Golam Abu Zakaria1, Wilhelm Schuette.
Abstract
For the determination of the absorbed dose to water for high-energy photon and electron beams the IAEA code of practice TRS-398 (2000) is applied internationally. In Germany, the German dosimetry protocol DIN 6800-2 (1997) is used. Recently, the DIN standard has been revised and published as Draft National Standard DIN 6800-2 (2006). It has adopted widely the methodology and dosimetric data of the code of practice. This paper compares these three dosimetry protocols systematically and identifies similarities as well as differences. The investigation was done with 6 and 18 MV photon as well as 5 to 21 MeV electron beams. While only cylindrical chambers were used for photon beams, measurements of electron beams were performed using cylindrical as well as plane-parallel chambers. The discrepancies in the determination of absorbed dose to water between the three protocols were 0.4% for photon beams and 1.5% for electron beams. Comparative measurements showed a deviation of less than 0.5% between our measurements following protocol DIN 6800-2 (2006) and TLD inter-comparison procedure in an external audit.Entities:
Keywords: Absorbed dose to water; DIN 6800-2 protocol (1997); DIN 6800-2 protocol (Draft National Standard 2006); IAEA TRS-398 protocol (2000); external beam photons and electrons
Year: 2007 PMID: 21217912 PMCID: PMC3003886 DOI: 10.4103/0971-6203.31143
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Characteristics of the ionization chambers
| Outer electr.-Ø | 5.5 mm | 6.1 mm | - | - |
| Inner electr.-Ø | 1.0 mm | 1.1 mm | - | - |
| Wall material | PMMA | PMMA + graphit | - | - |
| Wall thickness | 0.75 mm | 0.335 mm | - | - |
| Electrode spacing | - | - | 2.0 mm | 2.0 mm |
| Chamber-Ø | - | - | 6.0 mm | 15.0 mm |
| Membrane material | - | - | Polyaethylen | PMMA |
| Membrane thickness | - | - | 0.03 mm | 1.0 mm |
| Guard ring width | - | - | 4 mm | 0.2 mm |
| Cavity volume | 0.3 cm3 | 0.6 cm3 | 0.055 cm3 | 0.35 cm3 |
| Voltage | 400V | 400V | 300V | 100V |
Reference conditions for the determination of depth dose curve and absorbed dose to water in high-energy photon and electron beams (zref = reference depth) for all three protocols. IAEA recommends a field size of 10×10 cm2 for electron beams, but we have chosen a field size of 20×20 cm2 for all three protocols for comparison
| Field size at | ||||
| SSD: | 10×10 cm2 | 20×20 cm2 | 10×10 cm2 | 20×20 cm2 |
| SSD: | 100 cm | 100 cm | 100 cm | 100 cm |
| Measurement depth: | - | - | zref = 10 cm | zref = 0.6*R50 - 0.1 |
Possible ranges of application for different beam qualities and chambers in each dosimetry protocol
| Cylindrical chamber PTW-31013 ( | X | recommended ≥ 10 MeV | X | ≥ 10 MeV | X | recommended ≥10 MeV |
| Cylindrical chamber PTW-30006 ( | X | recommended ≥ 10 MeV | X | ≥ 10 MeV | X | recommended ≥ 10 MeV |
| Plane-parallel PTW -34001 ( | not re-commended | X | not re-commended | X | not re-commended | X |
| Plane-parallel PTW-23343 ( | not permitted | not permitted | not re-commended | X | not re-commended | not re-commended |
X = (here) unrestricted application
The reference conditions and the formalism of determination of the quality correction factors for photon beams kQ and electron beams kE(zref) in different protocols (kr = gradient correction factor, eff. point of measurement: the effective point of measurement is located at reference depth zref, ref. point of chamber axis: the effective point of measurement is located at zref + 0.5*rcav [rcav = inner radius of the sensitive volume])
| Cylindrical chamber | kQ° = kQ*kr | kQ° = kE'*kE”*kr | kQ° = kQ | kQ° = kQ | kQ° = kQ*kr | kQ° = kE' kE” *kr |
| kQ from | kE' = f (R50,Rp) kE” = pE/pCo | kQ from Table 14 | kQ from Table 18 | kQ from | kE' = f (R50) kE” from | |
| eff. point of measurement | eff. point of measurement | ref. point of chamber axis | eff. point of measurement | eff. point of measurement | eff. point of measurement | |
| Plane-parallel chamber | - | kQ° = kE'*kE” | - | kQ° = KQ | - | kQ° = kE'*kE” |
| - | kE' = f (R50,Rp) | - | kQ from Table 19 | - | kE' = f (R50) | |
| - | kE” from cross calibration ref. point of chamber axis | - | Nw from cross calibration ref. point of chamber axis | - | kE” from | |
Displacement from the chamber axis towards the radiation source (effective point of measurement) and displacement correction factor for the cylindrical chamber PTW 31013
| Photons | Displacement correction factor | kr = 1.007 | kr = 1.0084 | kr not used |
| Displacement from the | ||||
| chamber axis | 1.4 mm | 1.4 mm | 0 | |
| Electrons | Displacement correction factor | kr = 1.007 | kr = 1.0084 | kr in ko |
| Displacement from the chamber axis | 1.4 mm | 1.4 mm | 1.4 mm | |
Beam quality Q and beam quality correction factor kQ for different photon energies and protocols
| 6MV-Photons | ||||||
| Acc. 1 | 0.6747 | 0.9881 | 0.6751 | 0.9890 | 0.6751 | 0.9912 |
| 6MV-Photons | ||||||
| Acc. 2 | 0.6701 | 0.9887 | 0.6707 | 0.9896 | 0.6707 | 0.9918 |
| 18MV-Photons | ||||||
| Acc. 2 | 0.7743 | 0.9655 | 0.7732 | 0.9666 | 0.7732 | 0.9706 |
Influence quantities and their contributions to total uncertainties
| Nw | DIN 6800-2 (2006) | 0.45 | 0.45 | 0.45 |
| Depth of measurement | Estimation | 0.1 | 0.1 | 0.1 |
| FOA | Estimation | 0.1 | 0.1 | 0.1 |
| Leakage current | Manufacturer's figure | 0.2 | 0.2 | 0.2 |
| kP | DIN 6800-2 (2006) | 0.1 | 0.1 | 0.1 |
| ks | DIN 6800-2 (2006) | 0.1 | 0.1 | 0.1 |
| kρ | DIN 6800-2 (2006) | 0.17 | 0.17 | 0.17 |
| kQ or kE | IAEA TRS 398 | 1.0 | 0.9 | 0.6 |
| kQ or kE | DIN 6800-2 (2006) | 1.0 | 1.2 | 1.3 |
| Dosimeter reading | Manual PTW-UNIDOS | 0.5 | 0.5 | 0.5 |
| Long-term stability Dosimeter/year | Manual PTW-UNIDOS | 0.1 | 0.1 | 0.1 |
| Total uncertainty according to TRS 398 | 1.25 | 1.17 | 0.96 | |
| Total uncertainty according to DIN 6800-2 (2006) | 1.25 | 1.42 | 1.50 |
Beam quality R50 and beam quality correction factor kE for different electron energies and protocols
| 5MeV Acc. 1 | 1.85 | 0.9154 | 0.9348 | 0.9131 | 0.9375 | - | 1.0611 |
| 7MeV Acc. 1 | 2.58 | 0.9106 | 0.9266 | 0.9059 | 0.9267 | - | 1.0502 |
| 8MeV Acc. 1 | 3.04 | 0.9080 | 0.9219 | 0.9024 | 0.9212 | - | 1.0440 |
| 10MeV Acc. 1 | 3.92 | 0.9037 | 0.9140 | 0.8970 | 0.9120 | 0.9125 | 1.0335 |
| 12MeV Acc. 1 | 4.66 | 0.9005 | 0.9080 | 0.8933 | 0.9055 | 0.9095 | 1.0259 |
| 14MeV Acc. 1 | 5.26 | 0.8979 | 0.9032 | 0.8908 | 0.9008 | 0.9074 | 1.0205 |
| 6MeV Acc. 2 | 2.32 | 0.9123 | 0.9350 | 0.9083 | 0.9303 | - | 1.0539 |
| 9MeV Acc. 2 | 3.46 | 0.9059 | 0.9235 | 0.8996 | 0.9166 | - | 1.0388 |
| 12MeV Acc. 2 | 4.59 | 0.9007 | 0.9138 | 0.8937 | 0.9061 | 0.9098 | 1.0266 |
| 15 MeV Acc. 2 | 5.96 | 0.8943 | 0.9025 | 0.8882 | 0.8958 | 0.9051 | 1.0149 |
| 18 MeV Acc. 2 | 7.44 | 0.8894 | 0.8929 | 0.8834 | 0.8867 | 0.9012 | 1.0052 |
| 21 MeV Acc. 2 | 8.45 | 0.8866 | 0.8874 | 0.8805 | 0.8812 | 0.8992 | 1.0000 |
Figure 1Comparison of absorbed dose ratios for the three protocols in reference to DIN 6800-2 (2006) in photon beams from the linear accelerators ONCOR Impression (Acc. 1) and ONCOR Avant garde (Acc. 2)
Figure 3Comparison of absorbed dose ratios for the three protocols in reference to DIN 6800-2 (2006) as a function of electron energy (Linear accelerator Acc. 2)
Figure 2Comparison of absorbed dose ratios for the three protocols in reference to DIN 6800-2 (2006) as a function of electron energy (Linear accelerator Acc. 1)