Calibration of Elekta aSi EPIDs used as transit dosimeter.

S Cilla1, A Fidanzio, F Greco, D Sabatino, A Russo, L Gargiulo, L Azario, A Piermattei.   

Abstract

The transit in vivo dosimetry performed by the Electronic Portal Imaging Device (EPID), avoids the problem of solid-state detector positioning on the patient. Moreover, the dosimetric characterization of the recent Elekta aSi EPIDs in terms of signal stability and linearity enables these detectors adaptable for the transit in vivo dosimetry with 6, 10 and 15 MV photon beams. However, the implementation of the EPID transit dosimetry requires several measurements. Recently, the present authors have developed an in vivo dosimetry method for the 3D CRT based on correlation functions defined by the ratios between the transit signal, s(t) (w,L), by the EPID and the phantom mid-plane dose, D(m)(w,L), at the Source to Axis Distance (SAD) as a function of the phantom thickness, w, and the square field dimensions, L. When the phantom mid-plane was positioned at distance d from the SAD, the ratios st(w,L)/s't(d,w,L), were used to take into account the variation of the scattered photon contributions on the EPID as a function of, d and L. The aim of this paper was the implementation of a procedure that uses generalized correlation functions obtained by nine Elekta Precise linac beams. The procedure can be used by other Elekta Precise linacs equipped with the same aSi EPIDs assuring the stabilities of the beam output factors and the EPID signals. The calibration procedure of the aSi EPID here reported avoids measurements in solid water equivalent phantoms needed to implement the in vivo dosimetry method in the radiotherapy center. A tolerance level ranging between ±5% and ±6% (depending on the type of tumor) was estimated for the comparison between the reconstructed isocenter dose, D(iso) and the computed dose D(iso,TPS) by the treatment planning system (TPS).

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21214287     DOI: 10.7785/tcrt.2012.500178

Source DB:  PubMed          Journal:  Technol Cancer Res Treat        ISSN: 1533-0338


  2 in total

1.  A Simple Method for 2-D In Vivo Dosimetry by Portal Imaging.

Authors:  Stefano Peca; Derek Wilson Brown; Wendy Lani Smith
Journal:  Technol Cancer Res Treat       Date:  2017-06-06

2.  Evaluation of  interfraction setup variations for  postmastectomy radiation therapy using EPID-based in vivo dosimetry.

Authors:  Shengwei Kang; Jie Li; Jiabao Ma; Wei Zhang; Xiongfei Liao; Hou Qing; Tingqiang Tan; Xin Xin; Bin Tang; Angelo Piermattei; Lucia Clara Orlandini
Journal:  J Appl Clin Med Phys       Date:  2019-09-21       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.