Literature DB >> 21208240

Delayed ICD lead cardiac perforation: comparison of small versus standard-diameter leads implanted in a single center.

Roberto Rordorf1, Fabio Canevese, Alessandro Vicentini, Barbara Petracci, Simone Savastano, Antonio Sanzo, Edoardo Gandolfi, Roberto Dore, Maurizio Landolina.   

Abstract

BACKGROUND: An increased risk of delayed cardiac perforation (DCP) with active-fixation small-diameter ICD leads has recently been reported, especially with regard to the St. Jude Riata lead (St. Jude Medical, Sylmar, CA, USA). Few data on the risk of DCP in small versus standard-diameter leads implanted in a single high-volume center are available. Moreover, no data on the performances of St. Jude's new small-diameter Durata lead are as yet available. The aim of this study was to assess the incidence of DCP in small versus standard-diameter leads implanted in our center.
METHODS: Between January 2003 and October 2009, 437 small-diameter leads (190 Medtronic Sprint Fidelis [Medtronic Inc., Minneapolis, MN, USA], 196 Riata, 51 Durata) and 421 standard-diameter (>8 Fr) leads were implanted.
RESULTS: After a median follow-up of 421 days seven of 858 (0.8%) patients experienced DCP. The incidence of DCP was higher in patients with small-diameter leads than in those with standard-diameter leads (1.6% vs 0%, P = 0.01). No cases of DCP occurred among 371 passive-fixation leads versus 1.4% of events among active-fixation leads (P = 0.02). The incidence of DCP was 2.5% in Riata, 1% in Sprint Fidelis, 0% in Durata, and 0% in standard-diameter leads (P < 0.01 for Riata vs standard-diameter leads).
CONCLUSIONS: Small-diameter active-fixation ICD leads are at increased risk of DCP, a finding mostly due to the higher incidence of events in the Riata family. By contrast, passive-fixation small-diameter leads and standard-diameter leads appear to be safe enough regarding the risk of DCP. Our preliminary data suggest that the new Durata lead is not associated with an increased risk of DCP. ©2010, The Authors. Journal compilation ©2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21208240     DOI: 10.1111/j.1540-8159.2010.03002.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  10 in total

1.  Unexpected early complication of implantable-cardioverter defibrillator.

Authors:  Ahmed Abdelhafez; Nancy Wassef; Kai Hogrefe; Mohsin Farooq
Journal:  BMJ Case Rep       Date:  2018-06-06

2.  Late perforation of Durata ICD lead--a case report.

Authors:  Bharati Das; Sachin Yalagudri; S Maiya; Sunita Abhraham; J Kannan; Ravi Kishore
Journal:  Indian Heart J       Date:  2014 Jul-Aug

3.  Diaphragmatic Pacing as an Initial Presentation of Delayed Ventricular Lead Perforation.

Authors:  Luai Madanat; Kuldeep Shah; Richard Bloomingdale; Brian D Williamson
Journal:  J Innov Card Rhythm Manag       Date:  2022-05-15

4.  Delayed cardiac perforation of the Durata implantable cardioverter-defibrillator lead more than 1 year after implantation.

Authors:  Sandeep K Sharma; Jonathan W Weinsaft; James E Ip; Jim W Cheung
Journal:  HeartRhythm Case Rep       Date:  2016-06-16

5.  Missed diagnosis: Dual right ventricular lead perforation resulting in chest pain.

Authors:  Daniel W Nelson; Caroline Vloka; Jule Wetherbee
Journal:  HeartRhythm Case Rep       Date:  2016-10-03

Review 6.  Identification and management of right ventricular perforation using pacemaker and cardioverter-defibrillator leads: A case series and mini review.

Authors:  Mohammad Ali Akbarzadeh; Reza Mollazadeh; Salma Sefidbakht; Soraya Shahrzad; Negar Bahrololoumi Bafruee
Journal:  J Arrhythm       Date:  2016-06-30

7.  Serious case of the hiccups.

Authors:  Elizabeth V Saarel; Kevin Hinkle; Susan P Etheridge
Journal:  HeartRhythm Case Rep       Date:  2015-02-13

8.  Very late perforation by a Riata implantable cardioverter-defibrillator lead-A case report.

Authors:  Masatsugu Nozoe; Kazuo Sakamoto; Daisuke Nagatomo; Nobuhiro Suematsu; Toru Kubota; Masanori Okabe; Yusuke Yamamoto
Journal:  J Arrhythm       Date:  2018-03-30

9.  An unusual presentation of delayed lead perforation: It's never too late.

Authors:  Shashank Jain; Jude Clancy; Mark H Schoenfeld
Journal:  HeartRhythm Case Rep       Date:  2021-11-26

10.  Venous Stenosis After Transvenous Lead Placement: A Study of Outcomes and Risk Factors in 212 Consecutive Patients.

Authors:  Basil Abu-El-Haija; Prashant D Bhave; Dwayne N Campbell; Alexander Mazur; Denice M Hodgson-Zingman; Vlad Cotarlan; Michael C Giudici
Journal:  J Am Heart Assoc       Date:  2015-07-31       Impact factor: 5.501

  10 in total