Literature DB >> 21208230

Clinical predictors of cardiovascular implantable electronic device-related infective endocarditis.

Katherine Y LE1, Muhammad R Sohail, Paul A Friedman, Daniel Z Uslan, Stephen S Cha, David L Hayes, Walter R Wilson, James M Steckelberg, Larry M Baddour.   

Abstract

BACKGROUND: Cardiovascular implantable electronic device (CIED)-related infective endocarditis (CIED-IE) is a serious complication of cardiac device infection and is associated with increased mortality. At present, there exist no criteria to predict CIED-IE in patients who present with CIED infection.
METHODS: We retrospectively reviewed all cases of CIED infection seen at Mayo Clinic Rochester between 1991 and 2008. CIED-IE was classified using pathologic and clinical criteria. Clinical predictors of CIED-IE were identified using logistic regression, and quantified using a summary score and plotted against the distribution of CIED-IE.
RESULTS: Ninety-three (22.4%) of the 416 patients with CIED infection had CIED-IE. Host factors including chronic immunomodulator therapy exclusive of corticosteroid (odds ratio [OR], 3.79 [confidence interval (CI) 1.10, 13.04]), chronic corticosteroid therapy (OR, 2.15 [CI 0.93, 5.00]), hemodialysis (OR, 3.24 [CI 1.39, 7.55]), or remote infection (OR, 1.77 [CI 0.99, 3.14]) were associated with increased odds of CIED-IE. Patients with CIED-IE were at increased odds of presenting with fever (OR, 3.78 [CI 1.93, 7.40]), or malaise (OR, 1.87 [CI 1.02, 3.41]), and have findings of leukocytosis (OR, 3.61 [CI 1.51, 8.62]). In marked contrast, they were at decreased odds of exhibiting signs/symptoms of infection at the generator pocket site (OR, 0.19 [CI 0.10, 0.36]). Summary scores of 6 and 11 predicted CIED-IE in approximately 50% and 90% of cases, respectively.
CONCLUSIONS: Certain comorbid conditions and systemic manifestations of infection were associated with CIED-IE. In contrast, pocket site infection was negatively associated with CIED-IE. These findings should assist clinicians in identifying patients who would more likely benefit from further investigation of CIED-IE with transesophageal echocardiography. ©2010, The Authors. Journal compilation ©2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21208230     DOI: 10.1111/j.1540-8159.2010.02991.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  20 in total

1.  Cardiovascular implantable electronic device infections in left ventricular assist device recipients.

Authors:  Talha Riaz; Juhsien J C Nienaber; Larry M Baddour; Randall C Walker; Soon J Park; Muhammad Rizwan Sohail
Journal:  Pacing Clin Electrophysiol       Date:  2013-09-02       Impact factor: 1.976

2.  Impact of antiplatelet therapy on clinical manifestations and outcomes of cardiovascular infections.

Authors:  Ammar Habib; Larry M Baddour; M Rizwan Sohail
Journal:  Curr Infect Dis Rep       Date:  2013-08       Impact factor: 3.725

3.  Implantable electrophysiologic cardiac device infections: a risk factor analysis.

Authors:  D Raad; J Irani; E G Akl; S Choueiri; E Azar; J Abboud; C Afif
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2012-08-25       Impact factor: 3.267

4.  Clinical presentation and outcomes of cardiovascular implantable electronic device infections in hemodialysis patients.

Authors:  LaTonya J Hickson; Janelle Y Gooden; Katherine Y Le; Larry M Baddour; Paul A Friedman; David L Hayes; Walter R Wilson; James M Steckelberg; M Rizwan Sohail
Journal:  Am J Kidney Dis       Date:  2014-01-03       Impact factor: 8.860

5.  Management of Cardiac Electronic Device Infections: Challenges and Outcomes.

Authors:  Rikke Esberg Kirkfeldt; Jens Brock Johansen; Jens Cosedis Nielsen
Journal:  Arrhythm Electrophysiol Rev       Date:  2016

6.  Permanent pacemaker-associated actinomycetemcomitans endocarditis: A case report.

Authors:  Zhenhong Li; Jennifer Madeo; Shadab Ahmed; Alex Vidal; Amgad Makaryus; Jose Mejia; Tabassum Yasmin
Journal:  Germs       Date:  2013-09-01

Review 7.  Infective Endocarditis in the Elderly: Diagnostic and Treatment Options.

Authors:  M P Ursi; E Durante Mangoni; R Rajani; J Hancock; J B Chambers; B Prendergast
Journal:  Drugs Aging       Date:  2019-02       Impact factor: 3.923

8.  Cardiovascular Implantable Electronic Device Infections due to Propionibacterium Species.

Authors:  Abdelghani El Rafei; Daniel C Desimone; M Rizwan Sohail; Christopher V Desimone; James M Steckelberg; Walter R Wilson; Larry M Baddour
Journal:  Pacing Clin Electrophysiol       Date:  2016-04-22       Impact factor: 1.976

9.  Management of Cardiac Implantable Electronic Device Infection.

Authors:  Cristian Podoleanu; Jean-Claude Deharo
Journal:  Arrhythm Electrophysiol Rev       Date:  2014-11-29

10.  Clinical manifestations and management of left ventricular assist device-associated infections.

Authors:  Juhsien Jodi C Nienaber; Shimon Kusne; Talha Riaz; Randall C Walker; Larry M Baddour; Alan J Wright; Soon J Park; Holenarasipur R Vikram; Michael R Keating; Francisco A Arabia; Brian D Lahr; M Rizwan Sohail
Journal:  Clin Infect Dis       Date:  2013-08-13       Impact factor: 9.079