Bishnu Majumdar1, Narayan Prasad Patel, V Vijayan.
Abstract
The aim of this study is to derive the non-uniformity correction factor for the two therapy ionization chambers for the dose measurement near the brachytherapy source. The two ionization chambers of 0.6 cc and 0.1 cc volume were used. The measurement in air was performed for distances between 0.8 cm and 20 cm from the source in specially designed measurement jig. The non-uniformity correction factors were derived from the measured values. The experimentally derived factors were compared with the theoretically calculated non-uniformity correction factors and a close agreement was found between these two studies. The experimentally derived non-uniformity correction factor supports the anisotropic theory.Entities:
Keywords: Brachytherapy source; dosimetry; ionization chamber; non-uniformity correction factor
Year: 2006 PMID: 21206670 PMCID: PMC3004141 DOI: 10.4103/0971-6203.26893
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Measurement jig with 0.1 cc chamber and brachytherapy unit
Non-uniformity correction factor of the 0.1 and 0.6 cc ionization chamber for distances from 0.8 to 10.0 cm from theoretical and experimental study. Theoretical values were calculated by extrapolation method for the point source, whereas the measured values were for 0.35 cm length source
| 0.8 | 1.166 | 1.175 | - | - |
| 1.0 | 1.107 | 1.126 | 1.3207 | 1.3333 |
| 1.2 | 1.08 | 1.098 | 1.2403 | 1.2517 |
| 1.5 | 1.06 | 1.0681 | 1.176 | 1.1795 |
| 1.8 | 1.058 | 1.0495 | 1.123 | 1.1451 |
| 2.0 | 1.033 | 1.0422 | 1.116 | 1.1196 |
| 2.5 | 1.024 | 1.0304 | 1.0813 | 1.0836 |
| 3 | 1.019 | 1.022 | 1.0628 | 1.0595 |
| 4.0 | 1.012 | 1.0126 | 1.0404 | 1.0398 |
| 5.0 | 1.006 | 1.0099 | 1.033 | 1.0279 |
| 6.0 | 1.0045 | 1.0078 | 1.0248 | 1.0198 |
| 7.0 | - | - | 1.0131 | 1.0133 |
| 8.0 | 1.0044 | 1.005 | 1.0185 | 1.0138 |
| 10.0 | - | - | 1.011 | 1.0064 |
Figure 2Standard error vs. distance between source and chamber in the measurement of non-uniformity correction factor for 0.1 and 0.6 cc chambers (N = 10 to 25)