Small-field electron dosimetry for the Philips SL25 linear accelerator.

H Rashid1, M K Islam, H Gaballa, U F Rosenow, J Y Ting.   

Abstract

Electron-beam characteristics of a Philips SL25 linear accelerator have been studied. Central-axis percentage depth doses, cross-beam profiles and beam output factors of 6-, 10-, and 20-MeV beams, selected from the available energy range of 4 to 22 MeV, are reported in this paper. The main thrust of this work is to determine the systematic variation of beam characteristics, especially the output factor, with standard cone sizes and cerrobend beam-shaping cutouts down to a field size of 2 X 2 cm Output factors for the standard cones (open field) are energy dependent in a complex manner, increasing with the cone size for the 6-MeV beam whereas decreasing for 10- and 20-MeV beams. The output factor falls below unity at lower energies (6 and 10 MeV) for fields with at least one side smaller than 6 cm, and stays nearly constant for the 20-MeV beam. Measured output factors of small fields are least squares fitted by a second-order polynomial function. Output factors for small rectangular fields have been derived from the one-dimensional and square-root formulas, and the equivalent-square method. Only the one-dimensional formula predicts the measured output factors of highly elongated fields to within +/- 1% experimental uncertainties. Different cones with the same size electron cutout show a varied dose response, primarily due to variation in scattered electron contamination from the cones.

Mesh:

Year:  1990        PMID: 2120560     DOI: 10.1118/1.596471

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Comparison of two programs in calculating electron output factors at extended source-surface distances.

Authors:  Yunfei Hu; Jonathan Andrew Lambert; Yang Wang
Journal:  J Med Phys       Date:  2016 Jan-Mar

2.  Comparison between small radiation therapy electron beams collimated by Cerrobend and tubular applicators.

Authors:  Cristina Di Venanzio; Marco Marinelli; Alessia Tonnetti; Gianluca Verona-Rinati; Paolo Bagalà; Maria Daniela Falco; Antonio Stefano Guerra; Maria Pimpinella
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

3.  Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system.

Authors:  Imad Ali; Nesreen Alsbou; Salahuddin Ahmad
Journal:  J Appl Clin Med Phys       Date:  2021-11-25       Impact factor: 2.102

4.  A graphical user interface for an electron monitor unit calculator using a sector-integration algorithm and exponential curve-fitting method.

Authors:  James C L Chow; Grigor N Grigorov; Christopher MacGregor
Journal:  J Appl Clin Med Phys       Date:  2006-02-15       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.