Literature DB >> 21198468

Evaluation of a transportable capnometer for monitoring end-tidal carbon dioxide.

T Hildebrandt1, M Espelund, K S Olsen.   

Abstract

We compared a small and transportable Capnometer (EMMA™) with a reference capnometer, the Siesta i TS Anaesthesia. During air-breathing through a facemask, both the EMMA (nine modules) and reference capnometer sampled expired gas simultaneously. A wide range of end-tidal carbon dioxide values were obtained during inhalation of carbon dioxide and voluntary hyperventilation. The median IQR [range] difference between all sets of carbon dioxide values (EMMA - reference) was -0.3 (-0.6 to 0.0 [-1.7 to 1.6] kPa; n = 297) using new batteries, which was statistically significant (p = 0.04) and located to two of the nine EMMAs tested. Using batteries with reduced voltage did not influence the measurements. The 95% CI of the medians of the differences were -0.4 to -0.2. We conclude that the EMMA can slightly under-read the end-tidal carbon dioxide but is generally comparable with a free-standing monitor. The precision of the EMMAs was similar whether new batteries or batteries with reduced voltage were used.
© 2010 The Authors Anaesthesia © 2010 The Association of Anaesthetists of Great Britain and Ireland.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21198468     DOI: 10.1111/j.1365-2044.2010.06499.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  3 in total

Review 1.  Journal of Clinical Monitoring and Computing 2016 end of year summary: respiration.

Authors:  D S Karbing; S E Rees; M B Jaffe
Journal:  J Clin Monit Comput       Date:  2017-03-02       Impact factor: 2.502

2.  Comparison of end-tidal CO2 measured by transportable capnometer (EMMA™ capnograph) and arterial pCO2 in general anesthesia.

Authors:  Kyung Woo Kim; Hey Ran Choi; Si Ra Bang; Jeong-Wook Lee
Journal:  J Clin Monit Comput       Date:  2015-08-12       Impact factor: 2.502

3.  The Use of Catheter Mount Will Result in More Reliable Carbon Dioxide Monitoring under Fluid Exposing Conditions.

Authors:  Yongil Cho; Wonhee Kim; Tae Ho Lim; Hyuk Joong Choi; Jaehoon Oh; Bossng Kang; Youjin Kim; In Young Kim
Journal:  Emerg Med Int       Date:  2019-07-01       Impact factor: 1.112