| Literature DB >> 21195681 |
Abstract
The further development of derivatizing reagents for plasma amino acid quantification by tandem mass spectrometry is described. The succinimide ester of 4-methylpiperazineacetic acid (MPAS), the iTRAQ reagent, was systematically modified to improve tandem mass spectrometer (MS/MS) product ion intensity. 4-Methylpiperazinebutyryl succinimide (MPBS) and dimethylaminobutyryl succinimide (DMABS) afforded one to two orders of magnitude greater MS/MS product ion signal intensity than the MPAS derivative for simple amino acids. CD(3) analogues of the modified derivatizing reagents were evaluated for preparation of amino acid isotope-labelled quantifying standards. Acceptable accuracy and precision was obtained with d(3)-DMABS as the amino acid standards derivatizing reagent. The product ion spectra of the DMABS amino acid derivatives are diagnostic for structural isomers including valine/norvaline, alanine/sarcosine and leucine/isoleucine. Improved analytical sensitivity and specificity afforded by these derivatives may help to establish liquid chromatography tandem mass spectrometry (LC-MS/MS) with derivatization generated isotope-labelled standards a viable alternative to amino acids analysers.Entities:
Mesh:
Substances:
Year: 2010 PMID: 21195681 DOI: 10.1016/j.jchromb.2010.12.010
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205