Similarity of ST and T waveforms of 12-lead electrocardiogram acquired from different monitoring electrode positions.

Elin Trägårdh-Johansson1, Annika Welinder, Olle Pahlm.   

Abstract

BACKGROUND: It is not always feasible to use standard electrode placement for limb leads when recording the 12-lead electrocardiogram (ECG). Other electrode placements have been accepted during monitoring. Nonstandard electrode positions, however, fail to produce waveforms identical to those recorded from the distal limb positions that are standard for diagnostic interpretation. The purpose of the present study was to validate the ST-T-segment for an alternative "Lund system" of proximal limb electrode sites.
METHODS: Twelve-lead ECGs (standard, Mason-Likar, and Lund lead placement) were collected from 167 patients.
RESULTS: There were systematic differences between measurements from standard vs Mason-Likar, but not vs the Lund system. The 95% confidence intervals of measurement agreement were similar or less when comparing measurements from the Lund system vs the first standard recording with measurements for the 2 standard recordings.
CONCLUSION: The Lund system might constitute a uniform convention for "diagnostic" ECGs as well as for monitoring ECG applications with regard to ST-T waveforms.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21190700     DOI: 10.1016/j.jelectrocard.2010.11.012

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438


  3 in total

Review 1.  ECG monitoring leads and special leads.

Authors:  Johnson Francis
Journal:  Indian Pacing Electrophysiol J       Date:  2016-07-17

2.  Normal limits of ECG measurements related to atrial activity using a modified limb lead system.

Authors:  Jayaraman Sivaraman; Gandhi Uma; Sangareddi Venkatesan; Mangalanathan Umapathy; Vella Elumalai Dhandapani
Journal:  Anatol J Cardiol       Date:  2014-02-26       Impact factor: 1.596

3.  Modified limb lead ECG system effects on electrocardiographic wave amplitudes and frontal plane axis in sinus rhythm subjects.

Authors:  Sivaraman Jayaraman; Venkatesan Sangareddi; R Periyasamy; Justin Joseph; Ravi Marimuthu Shanmugam
Journal:  Anatol J Cardiol       Date:  2016-06-29       Impact factor: 1.596

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.