Literature DB >> 21189231

Interconnection of multichannel polyimide electrodes using anisotropic conductive films (ACFs) for biomedical applications.

Dong-Hyun Baek1, Ji Soo Park, Eun-Joong Lee, SuJung Shin, Jin-Hee Moon, James Jungho Pak, Sang-Hoon Lee.   

Abstract

In this paper, we propose a method for interconnecting soft polyimide (PI) electrodes using anisotropic conductive films (ACFs). Reliable and automated bonding was achieved through development of a desktop thermocompressive bonding device that could simultaneously deliver appropriate temperatures and pressures to the interconnection area. The bonding conditions were optimized by changing the bonding temperature and bonding pressure. The electrical properties were characterized by measuring the contact resistance of the ACF bonding area, yielding a measure that was used to optimize the applied pressure and temperature. The optimal conditions consisted of applying a pressure of 4 kg f/cm(2) and a temperature of 180 °C for 20 s. Although ACF base bonding is widely used in industry (e.g., liquid crystal display manufacturing), this study constitutes the first trial of a biomedical application. We performed a preliminary in vivo biocompatibility investigation of ACF bonded area. Using the optimized temperature and pressure conditions, we interconnected a 40-channel PI multielectrode device for measuring electroencephalography (EEG) signals from the skulls of mice. The electrical properties of electrode were characterized by measuring the impedance. Finally, EEG signals were measured from the mice skulls using the fabricated devices to investigate suitability for application to biomedical devices.
© 2011 IEEE

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21189231     DOI: 10.1109/TBME.2010.2102020

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  3 in total

1.  Solderable and electroplatable flexible electronic circuit on a porous stretchable elastomer.

Authors:  Gi Seok Jeong; Dong-Hyun Baek; Ha Chul Jung; Ji Hoon Song; Jin Hee Moon; Suck Won Hong; In Young Kim; Sang-Hoon Lee
Journal:  Nat Commun       Date:  2012       Impact factor: 14.919

2.  A multi-channel, flex-rigid ECoG microelectrode array for visual cortical interfacing.

Authors:  Elena Tolstosheeva; Víctor Gordillo-González; Volker Biefeld; Ludger Kempen; Sunita Mandon; Andreas K Kreiter; Walter Lang
Journal:  Sensors (Basel)       Date:  2015-01-06       Impact factor: 3.576

3.  Flexible Tricolor Flag-liked Microribbons Array with Enhanced Conductive Anisotropy and Multifunctionality.

Authors:  Qianli Ma; Wensheng Yu; Xiangting Dong; Ming Yang; Jinxian Wang; Guixia Liu
Journal:  Sci Rep       Date:  2015-09-28       Impact factor: 4.379

  3 in total