Literature DB >> 21187251

Complication of the use of a radiofrequency device in arthroscopic shoulder surgery: second-degree burn of the shoulder girdle.

Shalen N Kouk1, Bojan Zoric, William B Stetson.   

Abstract

Radiofrequency ablation (RF) energy devices are increasingly being used in arthroscopic surgery. This is a case report of a 53-year-old man who had second-degree burns of the shoulder and chest wall as a result of fluid overheating due to RF. During an extensive arthroscopic bursal dissection of the left shoulder, the suction device on the RF wand and the valve was left open, allowing fluid to drip onto the chest and arm of the patient. Three days after his surgery, on his first postoperative visit, the dressings were removed, and the patient was noted to have severe blistering on the lateral aspect of his arm and on the anterolateral aspect of his left chest wall. One year after the surgery, he still had residual scarring that was only mildly intermittently pruritic. The use of RF during arthroscopic surgery is very useful clinically; however, it is important to have a strict understanding of the potential hazards the thermal energy can cause to the surrounding soft tissue.
Copyright © 2011 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21187251     DOI: 10.1016/j.arthro.2010.10.016

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  6 in total

1.  Use of a hip arthroscopy flexible radiofrequency device for capsular release in frozen shoulder.

Authors:  Stephen R Thompson; Marie-Eve Lebel
Journal:  Arthrosc Tech       Date:  2012-05-11

Review 2.  Heat-Related Complications from Radiofrequency and Electrocautery Devices Used in Arthroscopic Surgery: A Systematic Review.

Authors:  Carola F van Eck; Tim A C van Meel; Michel P J van den Bekerom; Jacco A C Zijl; Bauke Kooistra
Journal:  Arthrosc Sports Med Rehabil       Date:  2021-02-23

3.  Keeping the shoulder arthroscopy patient dry.

Authors:  R Amr; J Relwani; E Tanton
Journal:  Ann R Coll Surg Engl       Date:  2011-11       Impact factor: 1.951

Review 4.  What are the current indications for use of radiofrequency devices in hip arthroscopy? A systematic review.

Authors:  Carlos Suarez-Ahedo; S Pavan Vemula; Christine E Stake; Zachary A Finley; Timothy J Martin; Chengcheng Gui; Benjamin G Domb
Journal:  J Hip Preserv Surg       Date:  2015-08-11

5.  The influence of different footprint preparation techniques on tissue regeneration in rotator cuff repair in an animal model.

Authors:  Andreas Ficklscherer; Michaela Serr; Thomas Loitsch; Thomas R Niethammer; Matthias Lahner; Matthias F Pietschmann; Peter E Müller
Journal:  Arch Med Sci       Date:  2016-06-14       Impact factor: 3.318

6.  The influence of radio frequency ablation on intra-articular fluid temperature in the ankle joint - a cadaver study.

Authors:  Philipp Ahrens; Dirk Mueller; Sebastian Siebenlist; Andreas Lenich; Ulrich Stoeckle; Gunther H Sandmann
Journal:  BMC Musculoskelet Disord       Date:  2018-11-24       Impact factor: 2.362

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.