Literature DB >> 21186483

[Target Z-axis optimized LASIK for 2 cases with decentered ablation].

Bing Liu1, Xiangrong Wang, Haixia Ru, Chunyan Yang.   

Abstract

PURPOSE: To report target z-axis optimized LASIK treating for two eyes from two patients with decentered ablation.
METHODS: The LaserSight AstraPro2.2 Z software is a topography-guided custom ellipsoid ablation platform. LASIK with target z-axis optimization was performed to restore pre-operative visual axis using a LaserSight SLX excimer laser (version 5.3,300Hz).
RESULTS: During preoperative examination, the uncorrected visual acuity (UCVA) was 0.5 in both eyes, and the best spectacle-corrected visual acuities(BSCVA) were 1.0 and 0.8. The decentered ablation zones were diagnosed with LaserSight AstraMax topography. Following target z-axis microtranslation, fifteen and eight μm of central corneal tissues were preserved in the two eyes, respectively. The UCVA was 1.2 in both eyes, and secondary ablation zones were both centered on the visual axis. Topography maps and UCVA were stable in both eyes at the time of final follow-up.
CONCLUSION: Using target z-axis optimized LASIK with the LaserSight AstraPro Planner 2.2 Z customized ablation software was an effective method to modify decentered ablation and restore the visual axis.

Entities:  

Mesh:

Year:  2010        PMID: 21186483     DOI: 10.3969/g.issn.1000-4432.2010.02.004

Source DB:  PubMed          Journal:  Yan Ke Xue Bao        ISSN: 1000-4432


  1 in total

1.  Using LaserSight Astrapro Planner 2.2 Z software in corneal topography-guided laser in situ keratomileusis for myopia with asymmetric corneal shape.

Authors:  Bing Liu; Wei Chen; De-Wang Shao; Hua Wang; Hai-Xia Ru; Min Zhang; Ying Wang; Chun-Yan Yang
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.