Active personal dosemeters in interventional radiology: tests in laboratory conditions and in hospitals.

I Clairand1, J-M Bordy, J Daures, J Debroas, M Denozière, L Donadille, M Ginjaume, C Itié, C Koukorava, S Krim, A-L Lebacq, P Martin, L Struelens, M Sans-Mercé, M Tosic, F Vanhavere.   

Abstract

The work package 3 of the ORAMED project, Collaborative Project (2008-11) supported by the European Commission within its seventh Framework Programme, is focused on the optimisation of the use of active personal dosemeters (APDs) in interventional radiology and cardiology (IR/IC). Indeed, a lack of appropriate APD devices is identified for these specific fields. Few devices can detect low-energy X rays (20-100 keV), and none of them are specifically designed for working in pulsed radiation fields. The work presented in this paper consists in studying the behaviour of some selected APDs deemed suitable for application in IR/IC. For this purpose, measurements under laboratory conditions, both with continuous and pulsed X-ray beams, and tests in real conditions on site in different European hospitals were performed. This study highlights the limitations of APDs for this application and the need of improving the APD technology so as to fulfil all needs in the IR/IC field.

Entities:  

Mesh:

Year:  2010        PMID: 21186215     DOI: 10.1093/rpd/ncq556

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Angular dependence of shielding effect of radiation protective eyewear for radiation protection of crystalline lens.

Authors:  Yuma Hirata; Toshioh Fujibuchi; Katsuya Fujita; Takayuki Igarashi; Eiji Nishimaru; Shogo Horita; Reiko Sakurai; Koji Ono
Journal:  Radiol Phys Technol       Date:  2019-10-15

2.  High filtration in interventional practices reduces patient radiation doses but not always scatter radiation doses.

Authors:  Roberto M Sanchez; Eliseo Vano; Pablo Salinas; Nieves Gonzalo; Javier Escaned; Jose M Fernández
Journal:  Br J Radiol       Date:  2020-11-24       Impact factor: 3.039

3.  Real-time, ray casting-based scatter dose estimation for c-arm x-ray system.

Authors:  Zaid Alnewaini; Eric Langer; Philipp Schaber; Matthias David; Dominik Kretz; Volker Steil; Jürgen Hesser
Journal:  J Appl Clin Med Phys       Date:  2017-01-24       Impact factor: 2.102

4.  Energy Transfer Assisted Fast X-ray Detection in Direct/Indirect Hybrid Perovskite Wafer.

Authors:  Lulu Liu; Weijun Li; Xiaopeng Feng; Chunjie Guo; Huimao Zhang; Haotong Wei; Bai Yang
Journal:  Adv Sci (Weinh)       Date:  2022-03-23       Impact factor: 17.521

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.