Literature DB >> 21185274

Proficiency testing outcomes of 3-hydroxyisovalerylcarnitine measurements by tandem mass spectrometry in newborn screening.

Timothy H Lim1, Víctor R De Jesús, Nancy K Meredith, Maya R Sternberg, Donald H Chace, Joanne V Mei, W Harry Hannon.   

Abstract

BACKGROUND: The use of tandem mass spectrometry (MS/MS) for the analysis of amino acids and acylcarnitines from dried-blood spots (DBS) has become routine practice in newborn screening laboratories. The Newborn Screening Quality Assurance Program (NSQAP) added 3-hydroxyisovalerylcarnitine (C5OH) into its routine quality control and proficiency testing (PT) DBS materials for MS/MS to assure the quality of C5OH screening. We report the results from NSQAP evaluations for C5OH-enriched DBS, and summarize participant screening practices based on their analytical methods.
METHODS: NSQAP prepared C5OH-enriched DBS materials for its participants. Laboratories reported quantitative and qualitative results. Bias plots of quantitative results were constructed using reported data and the results were sorted by an analytical method.
RESULTS: NSQAP participants reported PT specimen 3964 as outside of normal limits for C5OH. The mean C5OH value for derivatized and non-derivatized methods was 2.80 and 2.67 μmol/l, respectively. Reported data from other specimens showed a similar trend in derivatized vs. non-derivatized assay results. Differences in C5OH quantitative values were observed among laboratories using different internal standards.
CONCLUSIONS: C5OH MS/MS measurements in DBS assays varied by method and the choice of internal standards. The use of NSQAP's DBS materials allows harmonization of C5OH measurements by newborn screening laboratories worldwide. Published by Elsevier B.V.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21185274     DOI: 10.1016/j.cca.2010.12.021

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  3 in total

1.  Quantitative acylcarnitine determination by UHPLC-MS/MS--Going beyond tandem MS acylcarnitine "profiles".

Authors:  Paul E Minkler; Maria S K Stoll; Stephen T Ingalls; Janos Kerner; Charles L Hoppel
Journal:  Mol Genet Metab       Date:  2015-10-08       Impact factor: 4.797

2.  A verification of the application of the non-derivatized mass spectrometry method in newborns screening of metabolic disorders.

Authors:  Yulan Zheng; Yao Chen; Xiaolong Qiu; Weifen Chen; Qingying Lin; Yinglin Zeng; Hong Zhao; Wenbin Zhu
Journal:  Medicine (Baltimore)       Date:  2019-05       Impact factor: 1.817

3.  Harmonizing Newborn Screening Laboratory Proficiency Test Results Using the CDC NSQAP Reference Materials.

Authors:  Charles Austin Pickens; Maya Sternberg; Mary Seeterlin; Víctor R De Jesús; Mark Morrissey; Adrienne Manning; Sonal Bhakta; Patrice K Held; Joanne Mei; Carla Cuthbert; Konstantinos Petritis
Journal:  Int J Neonatal Screen       Date:  2020-09-17
  3 in total