S Farzad Mohammadi1, Arash Mazouri, Nazanin Rahman-A, Mahmoud Jabbarvand, Gholam A Peyman.
Abstract
UNLABELLED: We designed a globe-fixation system for use with animal eyes for surgical skills training. The system core is a cup with an adjustable aperture. Vacuum is exerted through the cup to capture the globe at the preequatorial region. The cup rides over a hollow base, sliding on its opening spurs. A ballast is screwed to the end of the thread of the cup to create tactile feedback and create a tendency to return to the primary position. The system provides optimum versatility for the practice of anterior segment procedures, namely stability, rotation, a tendency to return to the primary position, globe pressure adjustability, and a modifiable orbit size. FINANCIAL DISCLOSURE: Disclosures are found in the footnotes.Mesh:
Year: 2011 PMID: 21183095 DOI: 10.1016/j.jcrs.2010.10.026
Source DB: PubMed Journal: J Cataract Refract Surg ISSN: 0886-3350 Impact factor: 3.351