[Technical parameters to decrease the radiation dose from conventional and digital radiographs].

J N Foulquier1.   

Abstract

Reducing radiation dose while maintaining adequate image quality on conventional and digital radiographs requires optimal use of the unit. Additional filtration avoids unnecessary tissue exposure and improves photon transmission. Automatic exposure control may reduce exposure and dose. The volume of tissue imaged must be limited by the use of diaphragms and shutters or compression. Sensitive detectors with increased photon detection also contribute to reduce dose. Radiographic films combined to rare-earth screens also afford a good photon-conversion efficiency. Large area flat panel amorphous silicon x-ray sensors may also reduce dose up to 50% compared to films. Finally, calculation of the Kerma-area product independent of the source distance constitutes an important indicator of radiation dose.

Mesh:

Year:  2010        PMID: 21178896     DOI: 10.1016/s0221-0363(10)70178-5