Absence of acute adverse in-vitro effects on AS-1 RBCs and whole blood-derived platelets following prolonged exposure to 13.56 MHz radio energy.

Rodeina Davis1, Jerome Gottschall, Alfonso Gutierrez, Clive Hohberger, Dharmaraj Veeramani, Jerry Holcombe.   

Abstract

BACKGROUND: There is growing interest in radio frequency identification (RFID) technology application for tracking blood products to achieve higher productivity and safety in the transfusion medicine supply chain. We have conducted a limited study to assess the temperature and biological effects of 13.56 MHz RF radiation on RBCs and whole blood-derived platelets (WBDP) under extreme exposure conditions. STUDY DESIGN AND METHODS: Using an FDA-approved protocol, test units of both RBC and WBDP were subjected to approximately 100 watts of RF energy for an extended duration (23-25 h) to assess worst-case effects. Three replications of the test were performed.
RESULTS: Hemolysis after 23-25 hours of RF energy exposure was 0.09% and 0.05%, respectively, for TEST and CONTROL RBC units and well within the ≤1% limit in the FDA-approved acceptance criteria. For WBDP units, the mean pH of TEST and CONTROL units were 7.27 and 7.19, respectively, following 23-25 hours of RF energy exposure, and well above the ≥6.2 acceptance limit. Further, there was no detectable acceleration in cellular degradation of RBC and WBDP products. While there was minimal temperature rise, the relative temperature increase between TEST and CONTROL units never exceeded the 1.5°C acceptance criterion.
CONCLUSIONS: 13.56 MHz-based RFID technology is unlikely to have any significant temperature or biological effects on RBC and WBDP units under the normal operating conditions (a maximum of 4 watts RF power exposure for about 20 nonconsecutive minutes for RFID tracking during the life of the blood product).
© 2010 American Association of Blood Banks.

Entities:  

Mesh:

Year:  2010        PMID: 21175470      PMCID: PMC3051369          DOI: 10.1111/j.1537-2995.2010.02733.x

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  4 in total

1.  Adding RFID layer to blood safety loop.

Authors:  Karen Lusky
Journal:  CAP Today       Date:  2005-07

Review 2.  Radio frequency identification for prevention of bedside errors.

Authors:  Sunny Dzik
Journal:  Transfusion       Date:  2007-08       Impact factor: 3.157

3.  Radiofrequency identification technology can standardize and document blood collections and transfusions.

Authors:  S Gerald Sandler; Al Langeberg; Leo DeBandi; Joan Gibble; Charles Wilson; Charles L Feldman
Journal:  Transfusion       Date:  2007-05       Impact factor: 3.157

4.  Platelet surface P-selectin measurements in platelet preparations: an international collaborative study. Biomedical Excellence for Safer Transfusion (BEST) Working Party of the International Society of Blood Transfusion (ISBT).

Authors:  L J Dumont; T VandenBroeke; K A Ault
Journal:  Transfus Med Rev       Date:  1999-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.